B&K Precision 2712 User manual

OPERATING INSTRUCTIONS
MODEL 2712
DIGITAL MULTIMETER
75% relative humidity. Temperature Coefficient: 0.1 x (specified accuracy) per ℃. ( to 18, 28
to 50).
Altitude: 6561.7 feet (2000m). Power: Single standard 9-volt battery, NEDA
1604, JIS 006P, IEC 6F22.
Battery life: 150 hours typical with carbon-zinc. Dimensions: 165mm (H) x78mm (W) x42.5mm (D). Weight: Approx. 10.0 oz.(285g) including holster. Accessories: One pair test leads, one spare
fuse, 9V battery (installed), and Operating Instructions.
DC VOLTS
Ranges: 400mV, 4V, 40V, 400V, 1000V Resolution: 0.01mV Accuracy: ± ( 0.15% rdg + 10 dgts) on 400mV range
± (0.1% rdg + 5 dgts) on 4V to 1000V ranges
Input impedance: 400mV: >3MΩ; 4V ~ 1000V:2.3 Overload protection: 1000VDC or 750VAC rms
AC VOLTS (TRUE RMS) (45Hz – 1kHz)
Ranges: 400mV, 4V, 40V, 400V, 750V Resolution: 0.01mV Accuracy:
± (1.2% rdg + 20dgts) 45 ~ 60Hz on 400mV to 400V ranges ± (1.5% rdg + 20 dgts) 60 ~ 500Hz on 4V range ± (1.5% rdg + 20dgts) 60 ~ 1kHz on 40V to 400V ranges ± (2.0% rdg + 20 dgts) 45 ~ 500Hz on 750V range
AC+DC VOLTS (TRUE RMS) (45Hz - 1kHz)
Ranges: 400mV, 4V, 40V, 400V, 750V Resolution: 0.01mV Accuracy:
± (1.5% rdg + 20dgts) 45 ~ 60Hz on 4V to 400V ranges ± (2.0% rdg + 20 ± (2.0% rdg + 20dgts) 60 ~ 1kHz on 40V to 400V ranges
SAFETY INFORMATION
The following safety information must be observed to ensure maximum personal safety during the operation at this meter:
Test the meter on a known voltage before using it to determine if hazardous voltage is present.
Do not use the meter if the meter or test leads look damaged, or if you suspect that the meter is not operating properly.
Never ground yourself when taking electrical measurements. Do not touch exposed metal pipes, outlets, fixtures, etc., which might be at ground potential. Keep your body isolated from ground by using dry clothing, rubber shoes, rubber mats, or any approved insulating material.
Turn off power to the circuit under test before cutting, unsoldering, or breaking the circuit. Small amounts of current can be dangerous.
Use caution when working above 60V dc or 30V ac rms. Such voltages pose a shock hazard.
When using the probes, keep your fingers behind the finger guards on the probes.
Measuring voltage which exceeds the limits of the multimeter may damage the meter and expose the operator to a shock hazard. Always recognize the meter voltage limits as stated on the front of the meter.
SPECIFICATIONS
Display: 4¾ digit liquid crystal display (LCD) with a maximum reading of 40000. Analog bargraph: 40 segments with measurements 20 times per second. Polarity: Automatic, positive implied, negative polarity indication. Overrange: MSD(Most Significant Digit) blinks . Zero: Automatic. Low battery indication: The
displayed when the battery voltage drops below the operating level.
Measurement rate: 2 times per second, nominal. Auto power off: Approx. 30 minutes.
Operating environment: 0 to 50 at < 70% relative humidity. Storage temperature: -20 to 60 at < 80% relative humidity.
Find Quality Products Online at: sales@GlobalTestSupply.com
Accuracy: Stated accuracy at 23±5, <
is
"
"
www.GlobalTestSupply.com
± (2.0% rdg + 20 dgts) 45 ~ 500Hz on 750V range Crest factor: ≤3 AC coupled true rms specified from 10% to 100% of range. Input impedance: Overload protection: 1000VDC or 750VAC rms
CURRENT
Ranges: 40mA, 400mA, 10A Resolution: 1uA DC accuracy:
± (0.5% rdg + 10 dgts) on 40mA to 400mA ranges ± (1.5% rdg + 10 dgts) on 10A range AC accuracy: (TRUE RMS) (50Hz ~ 1kHz) ± (2.0% rdg + 30 dgts) on 40mA to 400mA ranges ± (2.5% rdg + 30 dgts) on 10A range
Crest factor: ≤3 AC coupled true rms specified from 10% to 100% of range. Voltage burden: 0.2V on 40mA, 10A ranges
Input protection: 0.5A/500V fast blow ceramic fuse
10A/500V fast blow ceramic fuse
10A Input: 10A for 60 seconds maximum followed by a 10 minute cooling period.
RESISTANCE
Ranges: 400Ω, 4kΩ, 40kΩ, 400kΩ, 4MΩ, 40MΩ Resolution: 0.0 Accuracy:
± (0.3% rdg + 15dgts) on 400Ω range ± (0.3% rdg + 5dgts) on 4kΩ to 400kΩ ranges ± (0.5% rdg +10 dgts) on 4MΩ range ± (1.5% rdg + 20 dgts) on 40MΩ range
Open circuit volts typical: 1.2Vdc (2.5Vdc on 40 Overl
oad protection: 500VDC or AC rms
CAPACITANCE
Ranges: 4nF, 40nF, 400nF, 4uF, 40uF Resolution: 1pF Accuracy: ± ( 3.0% rdg + 20 dgts) on 4nF range
± (3.0% rdg + 5 dgts) on 40nF to 400nF ranges ± (3.0% rdg + 5 dgts) on 4uF to 20uF ranges ± (5.0% rdg + 5 dgts) on 20uF to 40uF ranges
Overload protection: 500VDC or AC rms
FREQUENCY
Ranges: 100Hz, 1kHz, 10kHz, 100kHz, 500kHz Resolution: 0.01Hz Accuracy: ± ( 0.1% rdg + 10 dgts) Sensitivity: 5Hz ~ 100kHz: >500mV rms, 100kHz ~ 500kHz: >1.5V rms
dgts) 60 ~ 500Hz on 4V range
400mV: >3MΩ; 4V ~750V:2.3
1V on 400mA range
0Ω range)
2.Set the Function/Range switch to the desired voltage type (AC or DC) and