Anritsu ML2438A Data Sheet

Power Meters
ML2400A Series
Power meters that give you more.
Graphics Display. Up to 100k samples
per second produces accurate profiles of cellular, PCS, radar, aircraft control, and telemetry signals. Pulse top power is easily measured and averaged over single or multiple pulses.
Softkeys. Menu driven operation
simplifies test procedures
High Reliability. Rugged, splash resistant
chassis design handles drop shocks and rough field treatment. A front panel cover and softcase are optional for further environmental protection.
Peaking Meter. A dual bar
graph aids tuning procedures.
Trigger Key. Triggering controls apply high measurement speed
precisely when it’s needed. Perform on-pulse measurements over more than 80 dB dynamic range.
Accuracy. Excellent return loss sensor design
reduces typical mismatch uncertainty by half ­improving accuracy and repeatability. Less SWR ripple in measurement data improves production yield and quality control.
Sensor EEPROM. All MA2400A/B Series sensors
have built-in EEPROMs with factory Calibration Factor data. Six User Calibration Factor Tables allow extra frequency points or compensation for couplers and power attenuators.
2
A
NRITSU
P
OWER
M
ETERS
G
IVE
Y
OU
M
ORE
Min/Max. Monitor total
power excursion in either readout or graphical profiling mode displays.
ncrease settling speed
90 dB Sensor Range.
One MA2472A Power Sensor replaces two traditional 50 dB sensors and eliminates the need for switching and settling software in ATE systems.
Voltage. Fast voltage measurements
monitor bias conditions.
3
Voltage Input.
The analog input measures voltage or accepts the V/GHz signal from a synthesizer for automated sensor calibration factor correction.
TTL Trigger.
Precisely control peak power measurements and the start/stop timing of burst measurements.
Analog Outputs.
Dual analog outputs support corrected and scaled measurements or realtime dual channel output. Synthesizer interface controls include zero blanking.
Options 6 - 10.
Sensor and calibrator connectors (not shown in photo)can be located on the rear panel.
Printer.
Document signal performance quickly with the standard parallel printer output.
High Speed GPIB.
Accelerate ATE throughput with special “Fast mode”
Optional NiMH Battery
Provides 6 hours continuous operation.
ACCURACY, SPEED, FLEXIBILITY, ML2400A SERIES HAS IT ALL
ML2407/08A for N-CDMA and Narrow Pulse Power Measurements
ML2407A is the ideal solution for accurate measurements of IS-95 signals, giving average, peak and crest readings
Pulses down to 1 µs can be captured and displayed in graphical mode thanks to the
0.6 µs rise time of the MA2469A sensor. Test amplifiers response to narrow input pulses.
4
C
OMPREHENSIVE
R
ANGE
OFP
OWERSENSORS
Anritsu’s thermal sensors provide excellent power measurement accuracy over 50dB of dynamic range with more speed than any other thermal sensor available. Thermal sensors use Seebeck elements where the combined effect of a thermal gradient and charge migration between dissimilar metals gives a true reading of average power on any incident waveform. Anritsu thermal sensors have class leading SWR and built in EEPROM with calibration factor and linearity correction data. This results in assured accuracy when measuring any signal. Anritsu’s fast thermal power sensors improve sensor rise time and fall time to less than 4.0ms– an order of magnitude
The Anritsu MA2440A series high accuracy diode sensors have a built in 3 dB attenuator to minimize input SWR. They are used where the best measurement accuracy is required over
High Accuracy Diode Sensors
Standard Diode Sensors
Diode sensors have greater speed, sensitivity and dynamic range than thermal sensors. All Anritsu diode sensors use a dual diode architecture that gives improved sensitivity and dynamic range over single diode architectures. The MA2470A Series Power Sensors 90 dB dynamic range is both fast and accurate. Linearity is better than 1.8%, typically < 1.0% through 18 GHz. The ML2400A Series Power Meters allow test engineers to adjust low-level averaging controls for
Sensor EEPROM
All MA2400A/B Series Power Sensors through 50GHz store calibration data and model information within internal EEPROMS. The ATE system can automatically monitor system configuration - ensuring accurate, calibrated measurements.
Most ATE systems use multiple power sensors. When those sensors lack EEPROMs, the ATE system requires elaborate data entry screens and database management to store the sensor cal factor data. Further, operators must be trained to look for the annual calibration report -
rather than the potentially dated table on the sensor housing - before entering the data.
ATE designers no longer need to track sensor serial numbers and cal factor data
programmatically in the system control software. The
MA2400A/B Series’ internal
EEPROM has several benefits. Operator intervention is reduced. Significant ATE programming is eliminated. Accuracy and verifiable calibration configuration are assured.
Using the rear panel V/GHz input automates sensor calibration factor correction. The adjustable scale is compatible with most commercial synthesizers. Automatic calibration factor correction simplifies procedures and minimizes the opportunity for operator errors.
High Power Applications
Traditional high power sensors are expensive and have degraded accuracy specifications. Further, annual calibrations require more time and expense. Anritsu’s new User Calibration Factor Tables avoid these problems. Any attenuator or coupler can be compensated by entering frequency and attenuation values into the MA2400A/B Series Power Sensors
internal EEPROM. Six tables can be stored. The attenuation device can be semi-permanently attached; the power meter automatically applies compensation during the 0.0dBm, 50MHz calibration reference process. The User Calibration Factor Tables are easily deactivated - allowing the power sensor to be used stand-alone also.
faster than previous thermal sensors. Settled power measurements are now 10 times faster, that means reduced test time.
optimization of GPIB speed, sensor settling, and noise reduction – allowing the industry’s fastest -70 dBm settling time.
MA2470A power sensors offer an ideal combination of speed and dynamic range for general purpose power measurements. A single sensor replaces the two sensors that were previously required with sensors limited to 50dB dynamic range.
Fast Thermal Sensors
5
The new MA2480A series Universal Power Sensors will measure any modulated or multi-tone signal thanks to a patented sensor architecture with three diode pairs. Universal power sensors deliver over 80 dB of dynamic range with speed and accuracy. Average power measurements on WCDMA signals can now be made without the need for special power meters. Universal sensors are also ideal for power measurements on other digitally modulated carriers such
Universal Power Sensors
The MA2460A fast diode sensors from Anritsu have a rise time of 0.6µs. This together with a sensor video bandwidth of 1.25MHz makes them the ideal solution for power measurements on N­CDMA (IS-95) signals. The MA2460 sensors must be used with the ML2407/08A power meter. This combination of meter and sensor provides fast signal processing and sampling speeds. Average power,
Fast Diode Sensors
as HDTV, DAB or QAM modulated radio links.
The sensor architecture ensures that one of the diode pairs is always operating in its square law region. The meter selects the diode pair operating in its square law region and is designed so that even the peaks of CDMA signals are measured accurately. Anritsu’s three stage diode pair approach leads to a very much faster measurement time than the two stage approach used in previous generations of average power sensors. No slowing of measurement speed is observed at the switching points, making them transparent to the user.
Universal power sensors are also ideal for applications where multiple signals are present, such as intermodulation measurements and satellite multi carrier power loading measurements.
A unique additional capability of the Anritsu Universal power sensor is the ability to use it as a standard diode sensor for CW measurements. In this mode the fast response of diode sensors is maintained across the full dynamic range of the sensor, meaning that for the majority of users it is the only sensor that they will ever need – a truly Universal Power Sensor.
a large dynamic range, for example when measuring amplifiers. High accuracy diode sensors have a dynamic range of 87 dB compared to the 90 dB of standard diode sensors.
peak power and crest factor on N­CDMA signals can be measured and displayed. The MA2460 sensors are dual diode sensors and delivers over 80 dB of dynamic range, making them suitable for both open and closed loop power control testing. The sensors internal AC detection circuitry gives a guaranteed noise floor of –60 dBm with typical performance to –70dBm, even when measuring CDMA signals.
Pulses down to 1µs can also be captured and displayed thanks to the sensor rise time of 0.6µs. In profile mode the ML2407A meter can be used to measure average power across narrow pulses, an increasingly common test method for amplifiers in digitally modulated systems.
MA242XA Series Thermal Sensors
Traditional Thermal Sensor
1000
100
10
1
0.1
-70
-60
-50
-40
-30
-20
-10
0
10
20
Rdgs/sec
Pin
MA2470A, Sampling Rate MA2470A, Reading Rate
MA2470A, GPIB Continuous Auto-Averaged Readings
Traditional Diode Sensor
Traditional Thermal Sensor
100000
10000
1000
100
10
1
0.1
-70
-60
-50
-40
-30
-20
-10
0
10
20
Rdgs/sec
Pin
MA2470A Series Power Sensors.
90 dB dynamic range is combined with high speed. Rise time is < 4 µsec. Fall time is < 10 µsec, typical.
MA2420A/B Thermal Sensors.
Rise and fall times of less than 4 ms improve both settling time and measurement rate by more than two orders of magnitude over most of the useful dynamic range.
INPUT
DET B -ve
DET B +ve
DET A -ve
DET A +ve
6dB
Power
Divider
Detector
‘C’
11dB
Pad
6dB Power Divider
Detector
‘B’
17dB
Pad
Detector
‘A’
DET C -ve
DET C +ve
RF
Anritsu’s MA2480A series Universal Power Sensors use a unique architecture with three cascaded diode pairs, dividers and attenuators. One of the diode pairs is within its square law region, ensuring true average power measurements of any input signal.
Power meters are an essential instrument for the precise characterization of amplifier output power and gain. The Anritsu ML2438A is a true dual channel meter, two
independent signal channels eliminate multiplexing. Gain and output power are measured simultaneously. Fast responding thermal or diode sensors respond immediately to changes in power level to reduce total test time.
The combination of 90dB dynamic range sensors, fast settling time and versatile display settings make the ML2438A ideal for amplifier measurements. When an attenuator is required in front of the sensor, power offset tables can be entered into the sensors EEPROM. These tables apply frequency dependant offsets to the measured value, ensuring the best measurement accuracy at all times. When measuring the gain compression
The Anritsu ML2400A series of power meters has been designed to offer the best combination of speed, accuracy and flexibility in a low cost package.
From its Preset condition the ML2400A series delivers accurate power measurements with a minimum of key strokes. Integral sensor EEPROM’s store linearity and calibration factor data. This simplifies measurement set up and reduces human error. Basic power measurements can be made with just two operations; a sensor zero and entry of measurement frequency. Frequent recalibrations are eliminated as calibration data is memorized even after a power down cycle.
The Channel key gives immediate access to the instrument configuration for selection of one or two display channels, with different units or resolution for each. An analog bar graph simplifies manual tuning and peaking of power levels. Audible and visual alarms alert the user to failed measurements when testing against upper and lower limits. The Max/Min hold display feature is perfect for long term drift measurements or for quickly establishing the highest and lowest powers available from a test device.
Anritsu diode sensors deliver stable and repeatable measurements down to –70dBm, thanks to advanced signal processing with automated low level averaging and a very low meter noise floor.
6
D
ESIGNEDFOR
Y
OUR
A
PPLICATION
Flexibility and Convenience for standard power measurements
True dual channel display shows amplifier output power and gain simultaneously
Amplifier Gain Measurements
A unique feature of the ML2400A series is the ability to display pulsed power signals on a graphical display. Now it is easy to capture and display pulsed power or TDMA signals quickly and easily. In Profile mode the power meter has TTL or edge based trigger functions so that pulses can be captured and displayed in real time. Trigger delay and user settable data
Pulse profiling
Power vs Time mode provides a strip chart display of RF power variation.
Monitor a test device for gain and output power variation over time against; temperature, supply voltage or a component tolerance. Trouble shoot timing glitches or other intermittent power behaviour such as power control in a mobile telephone. Tune circuits for peak performance.
In power vs time mode the graphical display can represent the trace as average, maximum, minimum or both max and min. A time window of 1 minute to 24 hours can be set.
Power vs Time
Anritsu ML2438A
90 dB Sensors
UUT
V/GHz
P
O
50 dB Sensors
UUT
P
O
A A A
A
Switch
Switch
Thermal
Thermal
Diode
Diode
90 dB power sensors simplify test hardware and software. Accuracy is improved because mismatch uncertainty and insertion losses in the switches are eliminated. The dual channel ML2438A speeds execution time by eliminating test system software for multiple meter addresses, cal factor correction, switch control, gain ranging, and meter settling.
of an amplifier across a range of frequencies, the V/GHz input updates the calibration factor as the source frequency is changed. This reduces set up time during manual testing and eliminates the chance of errors.
capture timing provide graphical displays from 5µs to 999ms. Measure pulses, detect and avoid switching glitches, view the flatness of a pulse top – it is all possible with the ML2400A series profile mode.
For todays TDMA mobile communications systems such as GSM, PHS, PCS-1900, DCS-1800, DECT and IS-136 the ML2430A can capture and display individual time slots. Using frame triggering it is now possible to measure the average power across a time slot with greater accuracy
The Anritsu ML2400A series of power meters are designed to function with Anritsu 69A and 68B synthesized sweep generators to form an integrated test solution for swept power and frequency measurements. Finally, "real-time" tuning speed AND traceable power accuracy are available in a low cost instrument. Easily tune converters, mixers, and transceivers with power meter accuracy. Testing devices with differing input and output frequency is particularly convenient because power sensors are broadband devices. Total power is measured regardless of the device’s output frequency. By comparison, receiver based instruments require special calibrations and test set-ups.
production managers and operate with standard power meters.
CDMA systems can have spectral bandwidths of many megahertz. Standard diode sensors are not suitable for measurements on these systems as they give incorrect power measurements when used in their linear region (powers >-25dBm). Thermal sensors can be used as an alternative, but are too slow and limited in dynamic range for the demands of a production environment.
7
Frequency Sweep and Power Sweep
Modulated & Multi-tone signals
Source Sweep Graphic Display. Acquire power sweep or frequency sweep data at speeds greater than
fifteen sweeps per second (GPIB to PC screen update rate). The meter automatically generates sweep synchronization with Anritsu 68000B or 69000A series Synthesized Sweep Generators.
Examine GSM or other TDMA standards slot power. Advanced triggering modes allow frame triggering. Variable gate widths and trigger delay enable precise average power measurements in any single slot.
Anritsu’s new Universal power sensor has been specifically designed for the demands of wide bandwidth systems. Universal sensors have no bandwidth restrictions and do not rely on calculated power from sampled waveforms. Universal power sensors deliver 80dB dynamic range with all the speed and accuracy demanded by
than ever before, with better accuracy and traceability than is achievable with integrated radio test sets. A marker table displays average power between markers or absolute power at a marker position for simplified slot analysis.
Custom Mode gives a large numeric readout of the power in a pulse within a precisely defined capture window. The window is defined by positioning markers in the graphical display or direct entry of the pulses time data.
8
O
PTIMIZETEST
S
YSTEMPERFORMANCE
Lower calibration costs. Higher system throughput. Less programming time. Lower purchase price.
VXI and MMS Systems
The ML2400A Series power meters are faster than any VXI or MMS power meter. A single sensor accomplishes more types of measurements over more dynamic range than any offered for card based instrumentation. A single, inexpensive sensor handles peak and CW measurements over a 90 dB range. Thus, the system needs less switch control code, fewer sensors, and the whole system operates faster. Physically, the ML2400A Series occupies less space.
Triggering controls with wide internal trigger ranges eliminate the need for programmatic wait statements for switch and source transient settling times.
The power meter chassis is already RF shielded. That opens another slot in the EMI rack for other microwave equipment. The power sensor connectors are already on the front panel (or rear panel) where they’re needed, so there’s no need for special bulkhead adapters and extra cables.
Further, the ML2400A Series’ built-in display helps trouble-shoot problems quickly. There’s no need to press fifteen buttons just to see the power level on ‘Uplink # 4.’ The pulse profiling and synthesizer compatibility is built-in ­avoiding special software.
The power sensor’s internal EEPROM allows programmatic activation of User Calibration Factor Tables. This allows a single, standard power sensor to perform high power measurements with attenuators or couplers; no “special” sensors are required.
The obvious benefit of a fast power meter is improved test system throughput.
Less obvious are development costs and carrying costs. For example, many power meters claim high speed, but how many offer the advanced trigger controls to make that speed useful? Similarly, a meter lacking sensitivity requires excessive averaging at low power levels.
When an existing test system can’t be reprogrammed, the ML2400A Series Power Meters emulate older single or dual channel meters. Throughput improvement depends upon the number of power measurements and the minimal use of wait statements within the older code. Typically, test execution speed is two to ten times faster.
FUNCTION MATH
<>
>
>
DC V AC V
CHOICES
LEVEL
LOCAL
ENTER
Power
Off On
DC 1
AC 1
On / Off
Off On
LINE STBY
ON
FUNCTION DATA ENTRY POWER REF
SENSOR
1.00mW 50MHz DO NOT EX­CEED SENSOR MAX PWR
SPECIALCAL ON/OFF
ON/OFF
ON/OFF
RESOLN CAL PAC AUTO REG
STORE
SENSOR
RECALL
PWR REF EXIT
SHIFT ZERO OFFSET
DUTY CYCLE
dBm/W FREQ REL
SET RANGE
ENTER
PRESET
LOCAL
ZERO CAL FREQ REL
MENU
ESCAPE
dB/mWRECALL
LOCAL
E N T E R
CALIBRATOR
A
B
GPIB
CW MOD AVG PEAK PULSE AVG BURST AVG OFFSET FREQ CORR AVG CW MODAVG PEAK PULSE AVG BURST AVG OFFSET FREQ CORR AVG REMOTE SRQ TALK LISTEN
POWER I O
M O D E
M O D E
The ML2400A Series Power Meters combine the advantages of speed, accuracy, sensitivity, graphics display, voltage measurement and portability into a single power meter.
The MA2470A Series and MA2440A Series power sensors reduce maintenance and calibration costs. For a test system with only 50 dB range power sensors, twice the number of power sensors and four times the meters must be maintained. Switches and cables must be calibrated and their frequency responses logged into the test system software. Additional sensor inventory keeps the test system operating during calibration or repair intervals. Thus, costs are much lower with 90 dB range power sensors.
Anritsu Power Meters replace several types of specialty meters. Many companies use one type of meter for high speed, another for portability, another for low purchase price, and another for thermal sensor compatibility. Standardizing on a single power meter reduces capital and maintenance expense.
Switches
Sensors
Sensors
Meters
Meters
100%
75%
50%
25%
0%
90 dB
Sensors
50 dB
Sensors
Purchase Price
50 GHz Gain Measurement System
With 90 dB Dynamic Range
The ML2430A’s 90 dB range sensors reduce purchase costs and annual calibration expenses.
9
P
ORTABLE
A
ND
R
UGGEDFOR
F
IELD
U
SE
Toughened case. Battery operation. Remote Monitoring... Ideal for use in any location.
The ML2400A Series’ brings accuracy and convenience to field service applications.
For remote trouble-shooting, the full duplex RS-232 remote control automatically calls a pre-entered phone number whenever a limits threshold is exceeded. Just set the limit level, enter the phone number and connect a modem. The ML2400A Series continuously monitors power conditions and calls the host controller when a violation is detected.
The meter’s data acquisition settings can adjust to monitor average power or the burst power of specific timeslots. The RS­232 port uses the same commands as the GPIB. Contact your Anritsu representative for PC compatible software.
The optional NiMH battery with “Smart” technology provides 6 hours of continuous operation, accurate fuel gauging, and < 2 hour fast charge cycling. The accessory softcase protects the power meter and accessories from shocks and the weather.
All MA2400A/B power sensors have automated calibration factor correction and internal temperature compensation simplifies field procedures. Operators can focus on solving problems, rather than following lengthy test procedures.
Measurement procedures are further simplified through standardization. As the only power meter to combine fast sensors for profile displays and thermal sensors in one measuring system, service organizations can standardize on a single power meter.
Softkey menus simplify instrument control by making the user interface easier to understand. Status and
selection of meter settings are displayed on the menu as they are entered: hidden selections and special function keys are avoided. A convenient number entry pad avoids the procedural description problems associated with arrow keys used for numerical entries.
Anritsu Power Meters are designed for tough field handling and rugged conditions. The shock resistant polycarbonate case is thoroughly tested. Internal mountings for the battery and circuit boards are reinforced to survive ground impact. An optional front panel cover and softcase are recommended for further environmental protection.
Performance For Pulses
Individual time slots of GSM, DCS-1800, DECT, IS-136, PCS-1900 and PHS signals can be displayed and measured per the standards’ specifications.
Since measurements are performed with the same type of meter used by the equipment manufacturer, service personnel can easily correlate field installation and maintenance data to the manufacturer’s production tests. Problem conditions are easily identified.
Simplify Test Procedures
Remote Monitoring By Telephone
PC
Modem
Card
Public Switched
Telephone Network
Modem RS-232
ML2438A
Cell Site
10
Power measurement accuracy is a straight forward concept. It has several component parts. Some can be substantially reduced through judicious measurement practice. In the table below, a classic example of measurement accuracy is detailed for each MA2400A/B Series power sensor
type. A 16 GHz, 12.0 dBm signal is presumed at a source SWR of 1.5:1.0.
Instrumentation accuracy is the accuracy of the meter. The specification is 0.5% – a very small component of overall measurement accuracy. The error sources which comprise instrumentation accuracy are largely related to linear voltage measurement.
Sensor linearity and temperature linearity describe the relative power
level response over the sensor’s dynamic range. When measuring a power level at other than the power level of the absolute reference, which is typically the meter’s 0.0 dBm reference ‘calibrator’, sensor linearity is included
in measurement accuracy. Temperature linearity is included when operating the sensor at other than room temperatures. The best way to reduce this error is to first choose power sensors with clearly specified linearity performance.
Noise, Zero Set and Zero Drift impact measurement accuracy at the bottom of a power sensor’s dynamic range. Different power sensors have different noise characteristics. Noise can be reduced through additional averaging or longer sample integration times. Additionally, low level averaging can be applied to optimize tradeoffs between dynamic settling, raw noise power deviation, and measuring speed.
Mismatch uncertainty is typically the largest component of measurement uncertainty. The error is caused by the differing impedances between the power sensor and the device to which the power sensor is connected.The most convenient method of reducing mismatch uncertainty is to choose power sensors with high return loss; that is, the power sensor’s SWR is very close to 1.0. Further improvement results from performing actual measurements of the sensor’s return loss with a well calibrated vector network analyzer or other precision return loss measurement technique. Broadband microwave devices like the MA2400A/B Series sensors are assembled in a manner that typically achieves significantly better SWR performance at some frequencies than the performance specifications indicate.
There are three common techniques of reducing source SWR. If the source’s S22characteristics can be determined, the mismatch loss can be calculated and the appropriate correction added to
A
CCURACY
Sensor Model Series
Instrumentation Accuracy Sensor Linearity Noise, 256 Avg. Zero Set and Drift Mismatch Uncertainty Sensor Cal Factor Uncertainty Reference Power Uncertainty Reference to Sensor Mismatch Uncertainy Temperature Linearity, ± 20 C RSS, Room Temp Sum of Uncertainties, Room Temp RSS Sum of Uncertainties
MA2420A/B
0.50%
1.30%
0.00%
0.00%
3.67%
0.83%
1.20%
0.23%
1.00%
4.19%
7.73%
4.31%
8.73%
MA2440A
0.50%
1.80%
0.00%
0.00%
3.84%
0.79%
1.20%
0.23%
1.00%
4.51%
8.36%
4.62%
9.36%
MA2470A
0.50%
1.80%
0.00%
0.00%
4.49%
0.84%
1.20%
0.23%
1.00%
5.09%
9.06%
5.18%
10.06%
o
50.00
45.00
40.00
35.00
30.00
25.00
20.00
15.00
10.00
5.00
0.00
-70 -64 -58 -52 -46 -40 -34 -28 -22 -16 -10 -4 2 8 14 20
Typical 90 dB Sensor Accuracy
ML2470A Series Sensor Accuracy
% Uncertainty, RSS
Power, dBm
Typical Measurement Uncertainty for 90 dB Power Sensors
The MA2470A Series Power Sensors’ excellent return loss and low noise performance provide industry leading measurement accuracy. Chart shows typical 1.0 to 40 GHz performance excluding error sources for temperature, human error, and test setup frequency response. MA2440A Series High Accuracy Power Sensors further improve accuracy.
Mismatch is easily calculated in either dB or percentage terms from the source’s and sensor’s respective reflection coefficients.
The dual channel ML2438A provides realtime source leveling through a rear panel analog output and simultaneously measures RF power with the other sensor.
V/GHz to Meter
P
O
Reflection Coefficient = Γ
% Mismatch Uncertainty = 100 [( 1 ± Γ Γ ) -1 ]
dB Mismatch Uncertainty = 20 log ( 1 ± Γ Γ )
1
Reflection Coefficient = Γ
Ch 1: Analog Out
Power Sensor
2
2
1
2
1
ML2438A
2
Leveling In
To Other Analyzers
Amplifier
UUT
Filter
Optional
Attenuator
Limiter
11
the power measurement. Alternately, a
3.0 dB precision attenuator with high return loss can be attached to the source to improve the effective source impedance. Finally, effective source match can be improved with external power leveling.
Sensor calibration factor uncertainty identifies the accuracy of the sensor’s calibration relative to a recognized standard for absolute power level. Sensor calibration factor uncertainty is
included i n accuracy calculations for any absolute power measurement (in dBm or Watts) and for relative power measurements if the signals are different frequencies.
Reference power uncertainty specifies the maximum possible output drift of the power meter’s 50MHz, 0.0 dBm power reference between calibration intervals. The output is traditionally labeled ‘calibrator’ on a power meter’s front panel. Reference accuracy can be improved by using a more accurate or a more recently calibrated reference.
Relative power measurements (in dBr or %) can sometimes neglect several sources of error. Reference power uncertainty and sensor to reference mismatch uncertainty do not generally impact relative power measurements. If the measurements are performed at a single frequency, sensor calibration factor uncertainty is negligble. If the source or other device to which the power sensor is connected does not change its reflection characteristics in either magnitude or phase, then mismatch uncertainty can also be neglected under most conditions.
NIST, National Institute of Standards and Technology
Anritsu, Microwave Measurements Division Standards Labs
Anritsu Service Factory Customer Cal Lab
Power Meter End User
ML2430A Series measurements are NIST traceable.
In typical power measurements, the test setup and measurement practices often create larger measurement uncertainties. These errors include exceeding the sensors maximum specified power range, offset errors, poor connector practice, excessive source harmonics and incorrect calibration factor usage.
Offset errors occur because traditional power meters allowed only a fixed value power offset for test setup equipment. The frequency response characteristics from connectors, cables, adapters or attenuators, are assumed to be a single value. For example a ‘10 dB attenuator’ may only be 9.75 dB over some of the frequency range. The
0.25 dB error is approximately 6%, slightly larger than the total accuracy from other error components.
The ML2400A Series Power Meters avoid offset errors by incorporating up to five 200 point offset tables of dB values versus frequency. As frequency information is entered for sensor calibration factor correction, the correct offset level is also interpolated and applied. Compensating for the true frequency response of attenuators, couplers, cables, switches and other test setup devices improves measurement accuracy. When a power sensor connection is preceded with a 1N Series wideband power limiter, the offset table compensates for frequency response – achieving an accurate, “burnout-proof” sensor.
Connector damage has significant accuracy and repeatability effects. It is the most common cause of sensor damage, but is frequently undetected. Every MA2400A/B Series sensor includes a hex nut connection for application of a calibrated torque wrench. Historical common practice insists that “Finger tight is right.” For experienced operators, this may be acceptable. However, an RF connector is like any other precision tool; it has specifications. Particularly in production areas, quality practice dictates that no piece of measurement equipment be used outside of it’s tolerances. Torque wrenches assure compliance with that quality requirement and results in more consistent measurements.
ML2438A
Frequency Response
Stored to Table
Power Sensor
Attenuators, Couplers,
Cable, Limiters, Switches
dB
Frequency
Hex-nut connectors allow use of a calibrated connector torque wrench, assuring the connector is tightened to it’s proper tolerances.
All microwave devices have a frequency response. The ML2400A Series’ internal offset table automatically corrects for variation in test setup insertion loss versus frequency. The power sensor’s User Calibration Factor Tables accomplish the same function, except the data is stored within the sensor EEPROM rather than the meter.
12
A
CCESSORIES
Reduce annual calibration expense with Anritsu’s precision range calibrators. The ML2419A Range Calibrator verifies the ML2400A Series Power Meter’s measurement channels. The meter’s 50 MHz oscillator level is verified by comparison method. When the calibrator is connected, user operation prompts appear on the meter’s screen. A verification report is printed using the meter’s parallel printer port.
The DeskJet printer benefits from the print facility of the ML2400A series power meter to give immediate hard copy records of instrument settings, measurement conditions and results. Graphical displays are also output when the meter is in profile mode.
When power sensors must be located a long distance from the power meter, the MA2418A Reference Oscillator provides a remote, traceable 0dBm power reference. MA2418A is DC powered from a 16 to 24 volt jack plug.
When power sensor cables must pass through walls or shielded enclosures, the ML2400A-29 Bulkhead Adapter provides a convenient connection between two sensor cables. When connection to a 50MHz, 0dBm reference is required, the MA2418A Reference Oscillator provides a convenient solution.
MA2400A/B
Power Sensor
Wall
ML2400A-29
Bulkhead
Adapter
ML2438A
ML2419A Range Calibrator
50MHz Reference Oscillator
Sensor adapters protect your existing investment in power sensors. The MA2497A HP sensor adapter enables connection to HP 8480 series power sensors. MA2499B is for connection to the previous range of Anritsu 4700A and 4600A power sensors. Both sensor adapters have in-built EEPROMS so that calibration factor tables can be stored and recalled from the ML2430A power meter.
For field use the optional battery gives freedom of operation without access to a power supply. The battery can be charged in the instrument or with the accessory desk top charger. The accessory soft carry case has pockets for all common accessories, making it an ideal carry bag for field use.
The front panel cover protects the power meter display and connectors from knocks during transit. A bail arm is available for ease of carrying and angling the meter on a workbench.
Bulkhead Adapter
Soft Carry Case and Internal Battery
Bail Arm and Protective Front Cover
Printer
Anritsu and HP Sensor Adapters
13
Precision Coaxial Adapters
Model Frequency Range Connectors 510-90 DC - 3.3 GHz N male to 7/16 DIN female
510-91 DC - 3.3 GHz N female to 7/16 DIN female 510-92 DC - 3.3 GHz N male to 7/16 DIN male 510-93 DC - 3.3 GHz N female to 7/16 DIN male K220B DC - 40 GHz K male to K male K222B DC - 40 GHz K female to K female K224B DC - 40 GHz K male to K female
Coaxial Adapters
Model Frequency Range Connectors 1091-26 DC to 18 GHz N male to SMA male
1091-27 DC to 18 GHz N male to SMA female 1091-80 DC to 18 GHz N female to SMA male 1091-81 DC to 18 GHz N female to SMA female
Model Frequency Range Rating Connectors 1N50C 0.01 - 18 GHz 5W, 50 ohm N male to N female
1K50A 0.01 - 20 GHz 5W, 50 ohm K male to K female 1K50B 0.01 - 26 GHz 3W, 50 ohm K male to K female
Coaxial RF Limiters
Wideband limiters protect power sensors from damage due to excessive RF power. The limiters have low insertion loss to preserve the power sensor’s sensitivity. High quality return loss minimizes mismatch uncertainty degradation and ensures a flatter insertion loss versus frequency characteristic. The insertion loss response should be loaded into the MA2400A/B power sensor’s User Calibration Factor Table; however, it is important to note that the limiters large signal response, beginning at about + 10 dBm, is different from the limiter’s small signal frequency response.
Power Attenuators
Model Frequency Rating Connectors
Range
42N50-20 dc - 18 GHz 20 dB, 5W, 50 ohm N male to N female 42N50-30 dc - 18 GHz 30 dB, 50W, 50 ohm N male to N female 42KC-20 dc - 40 GHz 20 dB, 5W, 50 ohm K male to K female
RF Bridges and Open/Shorts
Model Description Frequency Range Connectors 60N50-1 RF Bridge, 46 dB 0.005 - 2.0 GHz N male
60NF50-1 RF Bridge, 46 dB 0.005 - 2.0 GHz N female 87A50-1 RF Bridge, 38 dB 2.0 - 18 GHz GPC-7 22A50 Open/Short DC - 18 GHz GPC-7 22N50 Open/Short DC - 18 GHz N male 22NF50 Open/Short DC - 18 GHz N female
Precision Loads
Model Frequency Range Connectors 28N50-2 DC - 18 GHz N male
28NF50-2 DC - 18 GHz N female 28A50-1 DC - 18 GHz GPC-7 28K50 DC - 40 GHz K male 28KF50 DC - 40 GHz K female 28V50B DC - 67 GHz V male 28VF50B DC - 67 GHz V female
Inexpensive RF limiters protect sensors against damage from excessive input power.
Precision Attenuators
Model Frequency Rating Connectors
Range
41KC-3 dc - 40 GHz 3 dB, 2W, 50 ohm K male to K female 41KC-6 dc - 40 GHz 6 dB, 2W, 50 ohm K male to K female 41KC-10 dc - 40 GHz 10 dB, 2W, 50 ohm K male to K female 41KC-20 dc - 40 GHz 20 dB, 2W, 50 ohm K male to K female 41V-3 dc - 60 GHz 3 dB, 2W, 50 ohm V male to V female 41V-6 dc - 60 GHz 6 dB, 2W, 50 ohm V male to V female 41V-10 dc - 60 GHz 10 dB, 2W, 50 ohm V male to V female 41V-20 dc - 60 GHz 20 dB, 2W, 50 ohm V male to V female
Power Dividers
Model Frequency Range Connectors 1091-29 DC - 18 GHz N male - N female/N female
K240B DC - 26.5 GHz K female - K female/K female K240C DC - 40 GHz K female - K female/K female V240C DC - 60 GHz V female - V female/V female
Power Splitters
Model Frequency Range Connectors 1091-28 DC - 18 GHz N female - N female/N female
K241B DC - 26.5 GHz K male - K female/K female K241C DC - 40 GHz K male - K female/K female V241C DC - 60 GHz V male - V female/V female
ML2400A-33 Desk-Jet Printer, Portable 2000-661 Black Print Cartridge 2000-662 Rechargeable Battery for Desk-Jet Printer 2000-663 Power Cable (Europe) for Desk-Jet Printer 2000-664 Power Cable (Australia) for Desk-Jet Printer 2000-665 Power Cable (U.K.) for Desk-Jet Printer 2000-667 Power Cable (So. Africa) for Desk-Jet Printer
Printers
Precision Waveguide to Coaxial Adapters
Contact your local Anritsu sales office for details of our range of precision waveguide to coaxial adapters.
Calibrated Torque Wrenches
Model Description 01-201 Calibrated torque wrench for K and V connectors
01-204 Calibrated torque wrench for N connector
14
Frequency Range:
100 kHz to 50 GHz (Sensor dependant)
Power Sensors:
Meter specifications apply to MA2400A/B Series Power Sensors. Compatible with MA and MP Series sensors.
Sensor Dynamic Range:
MA2420A/B Series Thermal Sensors: 50 dB
MA2440A Series High Accuracy Power Sensors: 87 dB CW, > 57 dB Peak
MA2460A/B Series Fast Diode Sensors: 80 dB
MA2470A Series Power Sensors: 90 dB CW, > 60 dB Peak
MA2480A Series Universal Sensors: 80 dB
Power Measurement Range:
-70 to +47 dBm (0.1nW to 50W), sensor/attenuator dependent. Use couplers for higher power levels.
Voltage Measurement Range:
0.00 to 20.00 V, nominal
Display Range:
-99.999 to +99.999 dB
Display Resolution:
Selectable from 0.1dB to 0.001dB limited to 0.01 dB in graphical display modes; Linear power units, 3 to 6 digit, 1 - 3 digits selectable to right of decimal nW
- W; Voltage, 1 - 2 digits selectable to right of decimal.
Offset Range:
-99.99 to +99.99 dB. Fixed value or frequency dependent table.
Display Units:
dBm, dB, dBr, dBmV, dBuV, W, %, Volts,
Instrumentation Accuracy:
< 0.5 %
Zero Set and Drift:
ML2437/8A < 0.5 % Percent of full scale in most sensitive range, measured over one hour with maximum averaging after one hour warm up at constant temperature.
ML2407/8A < 1.8 % Percent of full scale in most sensitive range, measured over one hour with maximum averaging after one hour warm up at constant temperature.
Noise:
ML2437/8A < 0.5 % Percent of full scale in most sensitive range,
ML2407/8A < 1.8 % Percent of full scale in most sensitive range, both measured over a one minute interval with maximum averaging, two standard deviations at constant temperature after one hour warm up, typical. MA2470A Series, 20 pW typical.
1.00 MW POWER REFERENCE Frequency:
50MHz nominal
Output Level:
1.00 mW, ±1.2%/Year, ± 0.9% RSS, NIST Traceable
Maximum Input:
+ 20 dBm continuous or peak, ± 50 V dc
VSWR:
< 1.04
Connector:
Type N female
SENSOR / CHANNEL CONTROL Operating modes:
Readout, dual channel. RF power or voltage. Power Versus Time: single channel graphic of
readout data Profile: single channel RF peak power graphic display
for analysis of repetitive pulse or transient waveforms
S
PECIFICATIONS
Resolution:
0.5 mV
Control:
Adjustable voltage to frequency Relationship
Analog Output (BNC):
Two outputs configurable to Log or Lin
Operating Modes:
Analog Out: Selectable channel adjusted for calibration factors and other power reading correction settings.
Pass/Fail: Selectable TTL High or Low Channel output: Near real time analog.
Uncalibrated. AC Modulation Output: Output 1 only. Dwell Output: Output 2 only
Output Range:
-5.0 to 5.0V
Resolution:
0.1 mV
Trigger Input: Operating Modes:
External TTL or RF Blanking.
GPIB Interface:
IEEE-488.2 and IEC-625
RS-232:
Supports software download and modem dial-out.
Parallel Printer Output:
Compatible with Deskjet 540 and 340 Models. Other 500 Series and 300 Series and later are typically compatible. Also Canon BJC 80. See manual for DIP switch settings.
GENERAL SPECIFICATIONS General:
MIL-T28800E,Type 3,class 5, Style E
Display:
Flat panel monochrome LCD graphic with backlight
Operating Temperature Range:
0.0 to +50 °C.
Storage Temperature Range:
-40 to +70 °C
Moisture:
Splash and rain resistant, 95% humidity non­condensing.
Power Requirements: AC:
90 to 250VAC,47 to 440Hz, 40VA Maximum
DC:
12 to 24 VDC, Reverse protected to -40V. Maximum input 30V.
Battery:
> 6 hr usable with 3000mAhr battery
Replaceable Battery(Option):
3000mAhr NiMH
EMI:
Complies with requirements for CE marking.
Warranty:
1 year standard, contact factory for extended warranty options
Dimensions:
8.39 inches (213mm) wide, 3.46 inches (88mm) high, 9.84 inches (390mm) deep
Weight:
<6.6Ibs (<3kg)
Source Sweep: Single channel power sweep or frequency sweep
NCDMA Average Power, Peak Power and Crest Factor-ML2407/8A only.
Range Hold:
Current range or selectable 1 through 5.
AVERAGING Auto-averaging:
Automatically increases moving averaging at low power ranges.
Averaging Types:
Auto, Manual (Moving, Repeat)
Manual Average Range:
1 to 512
Low-Level Averaging:
Low, Medium, and High settings apply post average low pass filter to improve visibility at high display resolution.
Limit Lines:
Fixed value high and low limits with audible, rear panel TTL output, and/or visible Pass/Fail alarm indication. Failure indication can latch for transient failure detection.
Cursors:
Two manually adjustable cursors with power, delta cursor power, between cursor power average, and delta time readout display.
Delta t Resolution:
0.5% of display period or 100ns
TRIGGERING Trigger Sources:
Internal, External TTL, GPIB, Manual, Continuous
Delay Range:
0.0 to 999.0 Milliseconds
Delay Resolution:
0.5% of display period or 100ns
Internal Trigger Range:
-15 to +20 dBm, all diode sensors. Selectable to -25dBm
Internal Trigger Level Accuracy:
1.0 dB, typical
External Trigger Range:
TTL rising or falling edge trigger. BNC input
Manual Trigger:
Front Panel Softkey
CHANNEL BANDWIDTH
ML2437/38A 100 kHz nominal ML2407/08A 1.4 MHz nominal
SYSTEM CONFIGURATION Display:
LCD Graphic display with backlight and adjustable contrast.
Save/Recall:
10 storage registers plus RESET default settings
Secure Mode:
Erases memory information upon power ON. Default condition is Secure Mode OFF.
Rear Panel Inputs/Outputs Cal Factor Voltage Input (BNC): Operating Modes:
Voltage: Display voltage reading on selected channel
Voltage proportional to frequency for sensor calibration factor compensation
Blanking Input: TTL levels only. Selectable positive or negative polarity.
Input Range:
0 to 20V
Standard Diode Sensors
15
Power Sensor Specifications
1. 0.0 dBm, room temperature.
2. Each MA2400A/B Series sensor incorporates precision RF connectors with hexagon coupling nut for attachment by industry standard torque wrench.
3. MA2460A/B Fast Diode Sensors must be used with ML2407/08A Power Meters for NCDMA and Fast Pulse measurements.
Model Frequency Range
Dynamic Range (dBm)
SWR Sensor Linearity
RF Conn
2
MA2472B 10 MHz - 18 GHz
- 70 to + 20
< 1.17; 10 - 150 MHz
MA2472B Only
< 1.90; 10 - 50 MHz < 1.17; 50 - 150 MHz < 1.12; 0.15 - 2 GHz < 1.22; 2 - 12.4 GHz < 1.25; 12.4 - 18 GHz < 1.35; 18 - 32 GHz < 1.50; 32 - 40 GHz < 1.63; 40 - 50 GHz
< 0.004
1.8%, < 18 GHz
2.5%, < 40 GHz
3.5%, < 50 GHz
N (m)
MA2473A 10 MHz - 32 GHz K (m)
MA2474A 10 MHz - 40 GHz K (m)
MA2475A 10 MHz - 50 GHz V (m)
MA2422B 10 MHz - 18 GHz
< 1.90; 10 - 50 MHz < 1.17; 50 - 150 MHz < 1.10; 0.15 - 2 GHz < 1.15; 2 - 12.4 GHz < 1.20; 12.4 - 18 GHz < 1.25; 18 - 32 GHz < 1.30; 32 - 40 GHz < 1.40; 40 - 50 GHz
N (m)
MA2423B 10 MHz - 32 GHz K (m)
MA2424B 10 MHz - 40 GHz K (m)
MA2425B 10 MHz - 50 GHz V (m)
MA2442B 10 MHz - 18 GHz
- 67 to + 20
< 1.17; 50 - 150MHz
MA2442B Only
< 1.90; 10 - 50MHz < 1.17; 50 - 150MHz < 1.08; 0.15 - 2GHz < 1.16; 2 - 12.4GHz < 1.21; 12.4 - 18GHz < 1.29; 18 - 32GHz < 1.44; 32 - 40GHz < 1.50; 40 - 50GHz
< 0.004
1.8%, < 18 GHz
2.5%, < 40 GHz
3.5%, < 50 GHz
N (m)
MA2444A 10 MHz - 40 GHz K (m)
MA2445A 10 MHz - 50 GHz V (m)
< 1.10; 0.1 MHz - 2 GHz < 1.15; 2 - 12.4 GHz < 1.20; 12.4 - 18 GHz
MA2421A 0.1 MHz - 18 GHz
- 30 to + 20
Fast Thermal Sensors
High Accuracy Diode Sensors
< 4.0
1.3%, < 18 GHz
1.5%, < 40 GHz
1.8%, < 50 GHz
N (m)
Universal Power Sensors
MA2481B 10 MHz - 6 GHz
-60 to + 20 N (m)
MA2482A 10 MHz - 18 GHz
MA2480/01 Adds fast CW mode to Universal Power Sensors for high speed measurements of CW signal plus TDMA and pulse measurements.
Fast Diode Sensors
MA2468A
3
10 MHz - 6 GHz
-60 to + 20 < 0.0006 1.8% N (m)
MA2469B
3
10 MHz - 18 GHz
< 1.90; 10 - 50 MHz < 1.17; 50 - 150 MHz < 1.12; 0.15 - 2 GHz < 1.22; 2 - 12.4 GHz < 1.25; 12.4 - 18 GHz
< 1.17; 10 - 150 MHz < 1.12; 0.15 - 2 GHz < 1.22; 2 - 12.4 GHz < 1.25; 12.4 - 18 GHz
10 MHz to 6 GHz
3% -60 to +20 dBm
6 to 18 GHz
3% -60 to 0 dBm
3.5% 0 to +20 dBm (1.8% CW with
option 1)
< 0.004
(with option
1 only)
Rise Time
1
(ms)
ANRITSU CORPORATION
5-10-27, Minamiazabu, Minato-ku, Tokyo 106, Japan Phone: +81-3-3446-1111 Telex: J34372 Fax: +81-3-3442-0235
Overseas Subsidiaries
·
U.S.A.
ANRITSU COMPANY
1155 E. Collins Blvd. Richardson, TX 75081, U.S.A. Phone: +1-800-ANRITSU Fax: +1-972-671-1877
·
Canada
ANRITSU Instruments Ltd.
4-205 Matheson Blvd. East, Mississauga Ontario, L4Z 3E3, Canada Phone: +1-905-890-7799 Fax: +1-905-890-2290
·
Brazil
ANRITSU ELECTRONICA LTDA.
Praia de Botafogo 440, Sala 2401 CEP 22250-040, Rio de Janeiro, RJ, Brasil Phone: +55-21-527-6922 Fax: +55-21-537-1456
·
U.K.
ANRITSU LTD.
200 Capability Green, Luton, Bedfordshire LU1 3LU, United Kingdom Phone: +44-1582-433200 Fax: +44-1582-731303
·
Germany
ANRITSU GmbH
Grafenberger Allee 54-56, D-40237 Düsseldorf 1, Germany Phone: +49-211-968550 Fax:+49- 211-9685555
·
France
ANRITSU SA
9, Avenue du Québec, ZA de Courtaboeuf 91951 Les Ulis Cedex, France Phone: +33-1-60-92-15-50 Fax: +33-1-64-46-10-65
·
Italy
ANRITSU SpA
Via Elio Vittorini, 129, 00144 Roma, Italy Phone: +39-6-509-9711 Fax: +39-6-502-2425 Fax: +968-791697
·
Sweden
ANRITSU AB
Botvid Center 145 84 Stockholm, Sweden Phone: +46-853470700 Fax: +46-853470730 Fax: +91-11-685-2275
·
Singapore
ANRITSU PTE LTD.
10, Hoe Chiang Road, #07-01/02 Keppel Towers, Singapore 089315 Phone: +65-282-2400 Fax: +65-282-2533
·
Hong Kong
ANRITSU COMPANY LTD.
Suite 812, 8/F, Chinachem Golden Plaza, 77 Mody Road, Tsimshatsui East, Kowloon, Hong Kong, China Phone: +852-2301-4980 Fax: +852-2301-3545
·
Korea
ANRITSU CORPORATION Limited
14F Hyunjuk Building, 832-41, Yeoksam­Dong, Kangnam-Ku, Seoul, Korea Phone: +82-2-782-7151 to 7156 Fax: +82-2-782-4590
·
Spain
ANRITSU ELECTRÓNICA S.A.
Europa Empresarial - Edifico Londres ­Planta 1. a- Oficina 6 Playa de Liencres, 2 28230 Las Rozas (Madrid) Spain Phone: +34 91 640 44 60 Fax: +34 91 640 44 61
Specifications are subject to change without notice.
15000-00004 rev C
Ordering Information
Models
ML2437A, Power Meter, single input ML2438A, Power Meter, dual input ML2407A, NCDMA Power Meter, single input ML2408A, NCDMA Power Meter, dual input
Included Accessories
Power cord for destination One 1.5m sensor cord per meter input Operation Manual Certificate of Calibration, also included with sensors.
Options and Accessories
ML2400A-01 Rack Mount, single unit ML2400A-03 Rack Mount, side-by-side ML2400A-05 Front Bail Handle ML2400A-06 Rear Mount Input A on ML2437A/07A ML2400A-07 Rear Input A and Reference on ML2437A/07A ML2400A-08 Rear Mount Inputs A, B and Reference ML2400A-09 Rear Mount Inputs A and B on ML2438A/08A ML2400A-11 Ni-MH Battery with Desktop Charger ML2400A-12 Front Panel Cover ML2400A-20 Spare1.5m Sensor Cable ML2400A-21 0.3m Sensor Cable ML2400A-22 3m Sensor Cable ML2400A-23 5m Sensor Cable ML2400A-24 10m Sensor Cable ML2400A-25 30m Sensor Cable ML2400A-26 50m Sensor Cable ML2400A-27 100m Sensor Cable ML2400A-29 Bulkhead Adapter ML2400A-30 Extra Operation Manual ML2437/8A ML2400A-31 Extra Operation/Programming Manual ML2407/8A ML2400A-33 Printer 2300-243 LabVIEW
Driver and Examples 760-209 Hardside Transit Case D41310 Soft Carry Case with shoulder strap 10585-00003 Maintenance Manual ML2400A Series MA2418A 50 MHz Reference Oscillator with Power Supply ML2419A Range Calibrator MA2499B Anritsu 4700 & 4600 Series Sensor Adapter MA2497A HP8480 Series Sensor Adapter
Options 1 to 5 above are mutually exclusive for any given ML2430A unit. Options 6,7, 8 and 9 above are mutually exclusive for any given ML2430A. Options 25, 26, 27 can not be used with ML2407/8A
Understanding SWR and Mismatch Uncertainty.
Return Loss, SWR, and reflection coefficient describe a device’s impedance match characteristic. When two devices of differing impedance are mated, energy from the output of the source device suffers incomplete transfer to the power sensor.
The partial reflections of microwave energy create a standing wave, hence the term standing wave ratio, SWR. A standing wave’s impact upon power measurement accuracy is analogous to moving the power sensor along the transmission line. The load within the power sensor detects more or less signal power depending upon these vector additions and subtractions. Thus, without knowledge of the vector reflection characteristics, it is not possible to know exactly where maxima and minima occur. The range of the minima and maxima are mismatch uncertainty.
In most test systems, the power sensor is attached at a fixed point on a transmission line. However, the effects of standing waves are still present because the position of the maxima and minima change as the frequency increases. At some point in a swept frequency measurement, in phase addition and subtraction is likely to occur. For this reason, mismatch uncertainty is calculated as a likely worst case value. When performing this calculation it is reasonable to neglect data sheet reflection specifications in favor of actually measuring the reflection magnitudes of the test components. Evaluating the actual reflection magnitudes within a test setup helps identify other causes of measurement variation. This practice is also a very effective means of identifying damage in coaxial connectors.
Standing waves are created when impedance mismatch causes RF signal reflections. The magnitude of the standing wave varies along the transmission line as the incident signal and reflected signal add and subtract vectorially.
The numerical values for reflection characteristics are mathematically related.
P
I
P
R
P
Power Sensor
MIN
P
Power Sensor
MAX
Distance From Source
Lower
Frequency
Frequency
P
I
F
H
Middle
F
L
F
M
High
Frequency
Power versus frequency graphs have amplitude ripple.
Reflection Coefficient = Γ =
VSWR -1
VSWR +1
= 10
P = P - P = P
D
RI ABSORBED
Power Sensor
Power Sensor Magnitude
Varies With Position
Power Sensor
- (Return Loss) / 20
> P
MIN
< P
MAX