Agilent M1165/66/67/75/76/77A
Component Monitoring System
and
Agilent M1205A V24 & V26
User’s Reference Manual
Volume 2
Parameter Information
s1
Part Number M1046-900 1L
Printed 06/2000
First Edition
Notice
This document contains proprietary information which is protected by copyright. All Rights
Reserved. Reproduction, adaptation, or translation without prior w rit ten permission is
prohibited, except as a llowed under the copyright laws.
Agilent Technologies
3000 Minuteman Road
Andover, MA 01810-1085
(978) 687-1501
Publication number
M1046-9001L
Warranty
The information contained in this document is subject to change without n otice.
Agilent Technologies ma kes no warranty of any kind with regard to this material, including,
but not limited to, the implied warranties or merchantability and fitness for a particular
purpose.
Agilent Technologies shall no t be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance, or use of this
material.
Copyright © Agilent Technologies, 2000
ii
Printing History
New editions of this document will incorporate all material updated since the previous edition.
Update packages may be issued bet ween editions and contain replac ement and additional
pages to be merged by a re vision date at the bottom of the pa ge. Note that pages which are
rearranged due to ch anges on a previous page are not considered revised.
The documentation printing date and part number indicate its current edition. The printing
date changes when a new edition is printed. (Minor corrections and updates which are
incorporated at reprint do not cause the date to change.) The document part number changes
when extensive technical ch anges are incorporated.
First Edition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . June 2000
Important
United States federal law restricts these devices to sale by or on the order of a physician.
The M1165/66/75/76A Systems comply with UL544, CSA 22.2-125, IEC 601-1, EN 60601-1, and
EN 60601-1-2 and car r ies Marking to Counci l Dir ec ti ve 93/ 4 2/ EEC, E ur opean Medical
Device Directive (MDD).
The M1167/77A Systems comply with UL2601-1, CSA 22.2 No. 601.1-M90, IEC 601-1,
EN 60601-1, and EN 60601-1-2 and carries Marking to Council Directive 93/42/EEC,
0366
European Medical Device Directive (MDD).
The M1205A Systems comply with UL2601, IEC 601-1, CSA C22.2 no. 601-1, EN60601-1, and
EN60601-1-2 and carries Marking to Council Directive 93/42/EEC, European Medical
Device Directive (MDD).
iii
Electromagnetic Interference
Anomalies due to electromagnetic interference are not unique to the M1165/66/67/75/76/77A
or the M1205A but are chara cteristic of patient monitors in use today. This performance is
due to the very sensitive high gain front end amplifiers used to display the physiological
signals. Among the many similarly performing monitors already in use by customers,
interference from electromagnetic sources is rarely a problem in actual use.
Avoiding Electromagnetic Interference
When electromagnetic interference (EMI) is encountered there are a number of actions that
can be taken to mitigate the problem.
• Eliminate the source. Possible sources of EMI can be turned off or moved away to
reduce their strength.
• Attenuate the coupling. If the coupling path is through the patient leads, the
interference may be reduced by moving and/or rearranging the leads. If the coupling is
through the power cord, plugging the monitor into a dif ferent circuit may help.
• Reduce the sensitivity of th e syst em . In all o f th e EM C t esti n g the m oni to r wa s adju ste d
to maximum sensitivity. For the ECG amplifier the gain was four times what is
normally required. By reducing the gain of the system receiving the EMI, the
interference can often be eliminated.
• Add external attenuat ors. I f E MI becomes an unusually difficult problem external
devices such as an isolation transf ormer or a transient suppressor ma y be of help. An
Agilent Customer Engineer can be of help in determining the need for external devices.
iv
Intended Use
Intended Use
Description
The Agilent M1165/66/67/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 Monitors are ne tw ork connectable bedside patient monitoring devices.
The Agilent M1205A Models V24CT and V26 CT may powered by either AC line power or by
battery power.
Purpose
The Agilent M1165/66/67/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 monitors measure and display multiple ph ysiological parameters and w aves, and
generate alarms and recordings. They exchange information with compatible devices. The
Agilent M1165/66/67 /75/76/77A Component Monitoring System and the Agile nt M1205A V24
and V26 monitors are not therapeutic devices.
Patient Population
The Agilent M1165/66/67/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 monitors are intended to be used on adult, pediatric, an d neonatal patients.
Environment
The Agilent M1165/66/67/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 monitors are intended to be used in a clinical environment by trained healthcare
professionals. They are not intended for home use.
They communicate with de vices such as a central station through network interface ports and
a serial I/O port.
The Agilent M1165/66/67/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 Monitors are pres cription devices and will carry the following label, “United
States Federal law restricts this device to sale by or on the order of a physic ian.”
v
Indications for Use
Indications for Use
Condition
The Agilent M1165/66/6 7/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 Monitors are generally indicated when the clinician decides there is a need to
measure and display multiple physiological parameters and waves, to generate alarms and
recordings of adult, pediatric, or neonatal patie nts.
Part of Body or Type of Tissue with Which the Device Interacts
The Agilent M1165/66/6 7/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 monitors do no t c ontact the body or tissue of the patient. Signals are obtain ed
from accessory electrode, transducer, and sensor devices.
Frequency of Use
The Agilent M1165/66/6 7/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 monitors ar e indicated for use when presc ribed by a clinician.
Physiological Purpose
The Agilent M1165/66/6 7/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 Monitors are indicated when the purpose is to gain information for treatment, to
assess adequacy of treatment, or to rule out causes of symptoms. The Agilent M1165/66/67/
75/76/77A Compon e nt Mo n it or i n g Syste m an d the Agi le n t M120 5A V24 an d V26 mo n itors are
well suited for patient m onitoring.
Patient Population
Adult, pediatric, and neonatal non-ambulatory patients.
vi
Indications for Use
Prescription Versus Over-the-Counter
The Agilent M1165/66/67/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 Monitors are prescription devices.
vii
Indications for Use
Warnings, Cautions, and Notes
Warnings, cautions, and notes are us ed throughout this User’s Manual to give you additional
information about the Agilent M1165/66/67/75/76/77A Component Monitoring System and the
Agilent M1205A V24 and V26 monitors. The warnings and cautions included in this safety
section refer to the equipment in general.
Wa r n i n g Warning
A “warning” calls attention to the user of imminent hazard to people if proper
procedures are not followed.
• For continued safe use of this equipment, it is necessary that the listed instructions are
followed. Instructions in this manual in no way supersede established medical
procedures.
• Explosion Hazard- Do n ot u se th i s equipment in the pre s en ce o f fl am ma bl e an esth etics.
• Alarms - Do not rely exclusively on the audible alarm system for patient monitoring.
Adjustment of alarm volume to a low level or off during patient monit oring may result
in patient jeopardy. Remember that the most relia ble method of patient monitoring
combines close persona l surveillance with correct operation of m onitoring equipment.
• This equipment is only intende d fo r use in healthcare facilities by trained healthcare
professionals.
• The product is not intended for outside hospital use such as a helicopt ers or
ambulances.
• This product is not intended for home use.
• To reduce the risk of elect ric al shock, do NOT remove any cover. Refer servicing to
qualified personnel.
• This equipment may interfere with ultrasound imaging equipment by causi ng
interference on the ultrasound display. Try to keep the instruments as far apart as
possible.
viii
Indications for Use
• Exposure of electrical contacts or connections to saline or other liquids and gels is
dangerous. Electrical contacts and connections such as cable connectors, power
supplies, parameter module plu g-in connections and rack connec tions must be kept
clean and dry. Thoroughly dry any electrical connections that become contaminated
with liquids. If additional decontamination is required please contact your biomedical
department or Agilent Tech nologies Response Center.
• Although this equipment is shielded against Electromagnetic Interference (EMI), it is
recommended to avoid the use of electrically radiating devices in close proximity to this
equipment.
• Connecting the Agilent monitoring network (SDN) cable when the product is powered
on is not supported. Error c odes and Agilent monitoring network (SDN) interface lockup may occur. Power cycling the product will recover the product. No permanent
damage will result. To prevent unintentional disruption in monitoring, be sure th e SDN
interface cable is properly secured at both ends when c onnecting to the Agilent
monitoring network (SDN).
• Do not connect a second rack by a cable when using a module rack docked to the back
of the Agilent V24CT or V26CT. Using a second rack connected by a cable may disrupt
module communication.
Caution
A “caution” calls attention to a c ondition or possible situation that co uld cause injury to the
user.
• Ventilation Requirements - Failure to meet ventilation requirements may cause
equipment failure an d, in turn, jeopardize the functions of automated monitoring. Do not
locate equipment in an enclosed ar ea which could restrict heat dissipation.
• Maintenance - Failure on the part of the responsible individual, hospital, or institution
employing the use of this equipment to implement a satisfactory maintenance schedule
may cause undue equipment failure and possible healt h hazards.
• Do not spray cleaning solut ions directly onto the monitor. Moisture droplets may enter
the internal components and cause equipment mal function or failure. Cleaning solutions
should be applied to a cloth and the cloth used to wipe the monitor clea n. The monitor
should be turned off during cleaning.
ix
Indications for Use
• Replacement Parts - It is highly recommended that only Agilent Technologies
recommended parts and a ccessories be used with this equipment. Failure to do so may
result in the degradation of performance. Accessorie s and parts for individual modules
and components are listed at the back of the appropriate section in th is manual.
Note—
A note gives special instructions to highlight an operating procedure or practice. Notes
may precede or follow the applicable text.
At this time, Agilent Technologies will ma ke available on request, and in E nglish only, such
circuit diagrams, component part lists, descriptions, calibration instructions, or other
information which wil l assist the user’s appropriate qualified technical pers onnel to repair
those parts of the equipment which are classified by Agilent Technologies to be repairable. A
list of Agilent Sales and Support Of fices is provided at the end of this manual.
x
Indications for Use
Using This Manual
To enable you to find information easily, there is a contents list at the front of the guide and a
comprehensive index at the back.
The User’s Reference Manual is separated into two parts; the core document and the
parameter module guides.
The Core Document
This section of the guide contains al l the general information about the sy stem. It is a good
place for new users to start because it gives an introd uction to the system and the way it
works, and shows you how to get start ed. Here is a list of the major sections:
• Introducing the Agilent M1165/66/67/75/76/77A Component Monitoring System and the
Agilent M1205A V24 and V2 6 monitors
• Getting Started
• Configuring the System
• Other Patients
• Alarm Functions
• Recording Functions
• Trends and Data Management
• Installation and Patient Safety
• Care and Cleaning
xi
Indications for Use
Parameter Module Sections
These sections each contain information for one parameter module. This covers setting-up,
monitoring, and problem solving if you encounter difficulty. Each section is separ at ed with a
white tab which has the title of the section.
Note—
The User’s Reference Manual contains information for all the parameter modules
available for the system. This means, of course, that depending on the model and number of
modules you have ordered, the screens w ill not always apply to your system. However, the
information for the param eters and functions is valid for all the system s.
Note—
The screenshots displayed in this manual were generated in demo mode and may
therefore differ from what actually appears on your screen during patient monitoring.
Notice to the User
Although there may be products in your area that look similar to the Agilent M1165/66/67/75/
76/77A Component Monitoring System and the Agilent M1205A V24 and V26 monitors, their
functionality may not be the same. This User’s Reference Manual is intende d t o be used with
the Agilent M1165 / 66/67/75/76/77A Component Monitoring System, the M1026A Anest hetic
Gas Module and the Agilent M1205A V24 and V26 monitors only.
This Manual is only applicable for Release C.0 versions of the monitors listed above. A
Release C.0 monitor can be identified by:
xii
a. the Release C.0 label on the monitor, or
b. the suffix of the EPROMpack part number. To view this number, press
→ → .
Monitor Revision Show SW Rev
The suffix of the EPROMpack part number on a Release C.0 Agilent CMS is A.
The Software R evision of a Releas e C.0 Agilent V24 or V26 monitor is L.xx.xx
Responsibility of the Manufacturer
Responsibility of the Manufacturer
Agilent Technologies only considers itself responsible for any effects on safety, reliability and
performance of the equipment if:
assembly operations, extensions, re-adjustments, modifications or repairs are carried out
by persons authorized by Agilent , and
the electrical installation of t he relevant room complies with nationa l standards, and
the instrument is used in accordance with the instructions for use.
To ensure optimum usage, we recommend that Agilent parts and accessories are used in
conjunction with the Agilent M1165/66/67/75/76/77A Co mponent Monitoring System, the
Agilent M1026A Anesthetic Gas Module and the Agilent M120 5A V24 and V26 Monitors,
wherever available. If non-Agilent parts are used, Agilent Technologies is not liable for any
damage that these parts may cause to the Agilent equipment.
Manufacturer´s Address
For South America, North America and Canada:
Agilent Technologies, Inc.
3000 Minuteman Road
Andover
MA 01810-1099
For all other countries:
Agilent Technologies GmbH
Herrenberger Str. 130
71034 Böblingen
Germany
xiii
Responsibility of the Manufacturer
xiv
Contents
Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
Printing History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Important . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Electromagnetic Interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Avoiding Electromagnetic Interference. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .v
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .v
Purpose. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .v
Patient Population . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .v
Environment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .v
Indications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
Condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
Part of Body or Type of Tissue with Which the Device Interacts. . . . . . . . . . . . . . . . . . . . . . vi
Frequency of Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
Physiological Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
Patient Population . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
Prescription Versus Over-the-Counter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Warnings, Cautions, and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii
Using This Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xi
The Core Document. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xi
Parameter Module Sections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xii
Notice to the User. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xii
Responsibility of the Manufacturer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiii
Manufacturer´s Address. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiii
ECG and ECG/Respiration Module Section 14-1
Introduction to ECG Parameter Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-2
What does it Measure?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-2
How the ECG/RESP Parameter Works . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-2
RESP Parameter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-6
Considerations When Monitoring ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-7
ECG Monitoring Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-9
Placing the Electrodes for ECG Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-10
Respiratory Monitoring Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-18
How Respiration is Measured . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-18
Contents-1
Checklist for Respiratory Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-18
Placing the Electrodes for Respiratory Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-19
ECG/RESP with Analog Respiration Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-20
ECG Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-21
Changing the Lead. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 4-21
Adjusting the Waveform Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-23
Changing the Filter Bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-25
Changing the QRS Detection Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-26
Adjusting the QRS Detection Level Manually . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-28
Correct Positioning of the QRS Detection Level. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 - 28
Changing Alarm Parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-29
Adjusting the Alarm Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-29
Switching ECG Parameter Alarms On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-29
Changing Brady Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-30
Tone Modulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-31
Multilead ECG Display and Recording. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-32
EASI™ 12-lead ECG Display, Printing and Recording . . . . . . . . . . . . . . . . . . . . . . . . . .14-33
Respiration Adjustments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-39
Changing the Respiration Detection Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-39
Adjusting the Respiration Detection Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-39
ECG and RESP Alarm and INOP Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-41
ECG Alarm Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-41
RESP Alarm Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-44
Parameter Settings Transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-46
Accessories and Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-47
ECG/RESP Performance Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-50
ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-50
RESP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-55
ECG/RESP with Analog Respiration Ou tput Performance Specifications
(CMS Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-56
Care and Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-57
ST Segment Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-58
ST Segment Monitoring with EASI™. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-58
Introduction to ST Segment Monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-59
Operation Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-60
Data Transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-61
ST Segment Monitoring Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 4-62
ST Segment Measure-ment Points. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-64
Waveform Storage/Recall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-65
Contents-2
Adjusting the Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-65
Adjusting the Alarm Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-66
ST Alarm and INOP Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-67
Parameter Settings Transfer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-70
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-71
ST Segment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-71
Noninvasive Blood Pressure Module Section 15-1
Introduction to the Noninvasive Blood Pressure Parameter Module. . . . . . . . . . . . . . . . . . . . . .15-2
NBP Measurement Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15-5
Notes on NBP Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15-7
Venous Puncture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15-11
NBP Setup Task Window. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15-11
Parameter Settings Transfer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15-13
NBP Alarm and INOP Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15-14
Measurement Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15-18
Accessories and Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15-19
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15-22
NBP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15-22
Care and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15-25
SpO2/PLETH Module Section 16-1
Introduction to the SpO2/PLETH Parameter Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-2
What does it Measure? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-2
How the SpO2/PLETH Parameter Works . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-3
Dual SpO2/PLETH Measure-ments (Agilent CMS only) . . . . . . . . . . . . . . . . . . . . . . . . . .16-5
SpO2/PLETH Measurement Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-6
Tone Modulation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-8
SpO2 Alarm Limits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-10
Escalating SpO2 Situations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-11
NBP Alarm Suppression. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-14
Testing the Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-14
Adjustment of the PLETH wave. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-15
SpO2/PLETH Alarm and INOP Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-20
Technical Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-22
Parameter Settings Transfer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-25
Measurement Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-26
SpO2/PLETH Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-28
Contents-3
SpO2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-28
PLETH. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-30
Care and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-30
Temperature Module Section 17-1
Introduction to the Temperature Parameter Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-2
Temperature Measurement Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17-4
Temperature Module Labeling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17-5
Labeling with Parameter Settings Transfer ON. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-6
Labeling with Parameter Settings Transfer OFF. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17-6
Temperature Labels Available. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17-7
Measuring Temperature Differences. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17-9
Getting into the Diff Temperature Task Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-10
Selecting Labels for Diff Temperatures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17-10
Parameter Settings Transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-11
Temperature Alarm and INOP Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17-12
Accessories and Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17-14
Performance Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17-16
Temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-16
Care and Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17-17
CO2 Module and Sidestream Module Section 18-1
Introduction to the CO2 and Sidestream Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18-2
Sidestream and Mainstream CO2 Measure-ment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18-2
CO2 Module Setup Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18-4
CO2 and Sidestream CO2 Measurement Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18-6
Mainstream Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18-8
Sidestream Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18-9
Sidestream Pump Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18-12
N2O, O2 and Environmental Corrections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18-13
Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18-16
Accuracy Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18-17
Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-18
CO2 Alarm and INOP Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18-21
Parameter Settings Transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-26
Accessories and Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18-27
Performance Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18-28
CO2/ssCO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18-28
Contents-4
Care and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18-33
FIO2 Module Section
(Agilent CMS only) 19-1
Introduction to the FIO2 Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19-2
What does it Measure? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19-2
How the FIO
FIO
Measurement Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19-4
2
FIO
Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19-7
2
Calibration in 21% O
Calibration in 100% O
Parameter Works. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19-2
2
2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19-7
2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19-8
FIO2 Alarm and INOP Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19-10
Parameter Settings Transfer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19-13
Accessories and Ordering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19-14
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19-15
FIO
Performance Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19-15
2
Care and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19-16
Pressure Module Section 20-1
Introduction to the Pressure Plug-in Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-2
Intra-Aortic Balloon Pump Pressure Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-3
Pressure Measurement Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-5
Zeroing the Transducer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-6
Changing the Wave Scale. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-8
Pressure Module Labeling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-9
Labeling with Parameter Settings Transfer ON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-10
Labeling with Parameter Settings Transfer OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-11
Pressure Labels Available. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-12
Pulmonary Artery Wedge Pressure Measurement (CMS only). . . . . . . . . . . . . . . . . . . . . . . . .20-14
Measuring Wedge Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-15
Starting the Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-17
Finishing the Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-18
Situations to Avoid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-19
Cerebral Perfusion Pressure Measurements (CMS only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-21
Application Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-21
Pressure Alarm and INOP Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-23
Flush/Sample Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-23
Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-23
Contents-5
Parameter Settings Transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-27
Accessories and Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-28
Performance Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-29
Invasive Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-29
Care and Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-32
Cardiac Output Module Section 21-1
Introduction to Cardiac Output Parameter Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21-2
Right Heart Thermo-
dilution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21-2
Transpulmonary
Thermodilution
(Agilent CMS only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-3
Cardiac Output Measurement - Right Heart Thermodilution . . . . . . . . . . . . . . . . . . . . . . . . . .21-5
Adjusting the Compu-tation Constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 1-8
Getting into the C.O. Measure-ment Task Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21-9
Measuring the Cardiac Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-10
Curve Alert Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-12
Task Window Prompt Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-13
Warning Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-13
Measuring the Blood Temp-erature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-14
Editing the Cardiac Output Measure-ment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-15
Cardiac Output Measurement - The PiCCO Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21-17
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21-17
Adjusting the Arterial Catheter Constant and the Injectate Volume . . . . . . . . . . . . . . . . .21-20
Injectate Temperature
and Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-21
Getting into the C.O. Measurement Task Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21-23
Measuring the Cardiac Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-24
Curve Alert Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-27
Task Window Prompt Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-28
Warning Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-28
Measuring the Blood Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21-29
Editing the Cardiac
Output
Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-30
Continuous Cardiac Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21-34
Contents-6
C.O. Alarm and INOP Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21-37
Errors in Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21-41
Parameter Settings Transfer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21-43
Accessories and Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21-44
Right Heart Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21-44
Transpulmonary Thermo
dilution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21-44
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21-46
Care and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21-49
VueLink Module Section 22-1
Introduction to the VueLink module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-2
Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-3
VueLink Module Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-6
Adjustments to VueLink Module Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-7
VueLink Module Task Windows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-7
VueLink Module Softkeys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-7
Selecting a Different Device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-8
General Wave and Numeric Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-10
Viewing Waves and Numerics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-12
Selecting Waves and Numerics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-15
Changing Wave Scale. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-17
VueLink Module Alarm and INOP Reporting from the External Device. . . . . . . . . . . . . . . . .22-19
Alarms Ignored. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-19
Alarms Accepted. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-19
Notes on Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-19
Notes on Inops . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-19
Alarm symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-20
VueLink Module Alarm and INOP Reporting from the System. . . . . . . . . . . . . . . . . . . . . . . .22-21
Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-21
INOP Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-21
SvO2 Module Section
(Agilent CMS only) 23-1
Introduction to the SvO2 Plug-in Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23-2
What does the Module Measure? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23-2
How Does the SvO
SvO
Main Task Window. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23-4
2
Measurement Work?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23-2
2
Contents-7
Setting up the SvO2 Equipment for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23-5
Data Storage and Recall in the Optical Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23-11
Alarms and INOP Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 3-12
Technical Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23-12
Parameter Settings Transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23-15
Accessories and Ordering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23-16
Performance Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23-17
SvO
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23-17
Care and Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23-18
2
Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23-18
Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23-18
tcpO2/tcpCO2 Module Section 24-1
Introduction to the tcpO2/tcpCO2 Parameter Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24-2
What does it Measure?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24-2
How the tcpO2/tcpCO2 Measurement Works . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24-2
Correlation of Transcutaneous with Arterial Blood Gas Values . . . . . . . . . . . . . . . . . . . . . 24-2
Setting up the Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24-3
Activating the Transducer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24-6
Preparing the transducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 4-7
1. Remembrane the tcpO2/tcpCO2 Transducer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24-7
2. Calibrating the tcpO2/tcpCO2 System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24-11
3. Applying the Transducer to the Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24-16
tcpO2/tcpCO2 Alarms and INOP Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24-20
Technical Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24-21
Parameter Settings Transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24-23
Accessories and Ordering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24-24
Performance Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24-25
tcpO2 and tcpCO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 4-25
Care and Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24-27
Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24-27
Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24-28
Disposal of Empty Calibration Gas Cylinders. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24-29
Ventilator Interfaces and Respiratory Loops (Agilent CMS only) 25-1
Ohmeda Ventilator Interface Section - 7800/7810 25-3
Introduction to the 7800/7810 Ohmeda Ventilator Interface . . . . . . . . . . . . . . . . . . . . . . . . . . .25-4
Contents-8
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-5
Ohmeda Ventilator Parameters and Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-6
Operating Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-7
Ohmeda Ventilator Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-7
Ohmeda Ventilator Signals and Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-8
Ohmeda Ventilator Alarm Reporting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-10
Alarms Accepted. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-10
Alarms Ignored. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-11
Ohmeda Ventilator Alarm and INOP Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-12
Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-12
INOP Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-12
Ohmeda Ventilator Interface Section - 7900 25-15
Introduction to the 7900 Ohmeda Ventilator Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-16
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-17
Ohmeda Ventilator Parameters and Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-18
Operating Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-19
Ohmeda Ventilator Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-19
Ohmeda Ventilator Signals and Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-20
Ohmeda Ventilator Alarm Reporting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-23
Respiratory Loops 25-25
Introduction to Respiratory Loops. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-26
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-27
Respiratory Loops Parameters and Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-28
Loop Types. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-29
Color Coding of Loops. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-30
Operating Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25-30
Anesthetic Gas Module Section 26-1
Section I:
The Anesthetic Gas Module – Option A01 26-3
Introduction to the Anesthetic Gas Module (Nafion Version). . . . . . . . . . . . . . . . . . . . . . . . . . .26-4
What does it measure? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-4
How the Gas Measure-ment Works . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-5
How fluids are removed from the gas sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-5
Front Panel of the Anesthetic Gas Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-6
Contents-9
Rear Panel of the Anesthetic Gas Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-8
Anesthetic Gas Module Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-9
What you need . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-9
Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-10
Detecting and Preventing Leaks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-16
Anesthetic Gas Exhaust. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-17
What you need to return the gas sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-17
Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-18
What you need to remove the gas sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-19
Anesthetic Gas Module Standby Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-20
Automatic Switch to Standby Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-21
Using the Anesthetic Gas Module during a Cardiopulmonary Bypass. . . . . . . . . . . . . . . . . . .26-22
Screen Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-23
Main Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-23
Automatic Parameter Switch Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-23
The Overview Task Window. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-24
The Airway Gas Task Windows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-25
Automatic Agent Identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-26
Selecting an Anesthetic Agent. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-26
Exchanging Agents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-27
Oxygen (O2) Measurement (Optional). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-29
Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-30
Zero Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-30
Span Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-33
Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-34
Accessories and Ordering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-42
Performance Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-42
Anesthetic Gas Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-42
Care and Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-48
Anesthetic Gas Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-48
Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-48
Section II:
The Anesthetic Gas Module – Option A02 & A05 26-55
Introduction to the Anesthetic Gas Module (Watertrap Version). . . . . . . . . . . . . . . . . . . . . . .26-56
What does it measure? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-56
How the Gas Measurement Works . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-57
How fluids are removed from the gas sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-57
Contents-10
Front Panel of the Anesthetic Gas Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-58
Rear Panel of the Anesthetic Gas Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-60
Anesthetic Gas Module Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-61
What you need . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-61
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-63
Detecting and Preventing Leaks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-67
Anesthetic Gas Exhaust . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-68
What you need to return the gas sample. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-68
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-69
What you need to remove the gas sample. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-70
Anesthetic Gas Module Standby Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-71
Automatic Switch to Standby Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-72
Using the Anesthetic Gas Module during a Cardiopulmonary Bypass . . . . . . . . . . . . . . . . . . .26-73
Screen Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-74
Main Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-74
Automatic Parameter Switch Off. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-74
The Overview Task Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-75
The Airway Gas Task Windows. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-76
Automatic Agent Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-77
Selecting an Anesthetic Agent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-78
Exchanging Agents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-78
Emergence from Anesthesia. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-79
Oxygen (O2) Measurement (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-80
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-81
Zero Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-81
Span Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-84
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-85
Accessories and Ordering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-93
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-94
Anesthetic Gas Module Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-94
Care and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-101
Anesthetic Gas Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-101
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26-101
Blood Analysis 27-1
Introduction to the Blood Analysis Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-2
Blood Analysis Module Keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-3
Obtaining Blood Specimens. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-5
Suitable Specimens. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-5
Contents-11
Criteria For Specimen Rejection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-5
Circum-stances to Avoid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 7-6
Procedure for Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-7
Preparing Cartridges for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-7
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-7
Blood Analysis Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-9
Selecting an Operator ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-10
Selecting the Blood Sample Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-11
Selecting a Patient Tem-perature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-11
Entering an O2 Value. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-11
Entering Information into the Free-Entry Fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-12
Blood Analysis Results Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-12
Blood Analysis Results. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-14
Transmitting Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-15
Procedure for Trans-mitting Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-17
Recording Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-17
Alternative Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-18
Blood Analysis Alarm and INOP Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-19
Technical Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-19
Accessory Ordering Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-20
Cartridges. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-20
System Control and Verification Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-21
Care and Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-22
EEG Module Section (Agilent CMS only) 28-1
Introduction to EEG Parameter Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-2
How the EEG Parameter Works . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-3
Considerations When Monitoring EEG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-5
EEG Monitoring Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-6
Placing the Electrodes for EEG Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-7
EEG Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-11
Changing the Scale. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28-12
Selecting the Numerics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28-13
Changing the CSA Epoch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-15
Low & High Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-16
Changing the Impedance Range Limit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-17
Print-Out. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-19
Switching CSA Screen On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-22
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-22
Contents-12
EEG INOP Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-23
Accessories and Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-25
Care and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-26
EEG Performance Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-27
A. Summary of Formulas Used in Calculations A-1
Hemodynamics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Oxygenation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-6
Abbreviations and Unit Definitions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-7
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-12
B. Analog Output Section
(Agilent CMS Only) B-1
Introduction to the Analog Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
Technical Terms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
Gain. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
Application Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-4
Scaled Waves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-4
Absolute Waves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-4
Numeric Trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B- 4
Configuring Scaled Wave Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5
Selecting the channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5
Assigning the\Scaled Wave . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6
Adjusting the Gain and Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-7
Configuring Absolute Wave Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-9
Selecting the Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-9
Assigning the Absolute Wave. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-10
Adjusting the Gain and Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-11
Configuring Numeric Trend Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-12
Selecting the Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-12
Assigning the Numeric Trend. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-13
Adjusting the Gain and Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-14
C. Calibrating the Pressure System C-1
Methods of Calibrating the Pressure System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
Zeroing the Transducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
Contents-13
Procedure for Zeroing the Transducer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-4
Mercury Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-6
Procedure for Mercury Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-8
Mercury Calibration Task Window. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-9
D. SpO2 Transducer Information D-1
Description of Transducers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-2
Transducers and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-3
Transducer Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-5
Application Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-6
Care and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-14
E. Supported Device Information E-1
F. Main Sales and Support Offices F-1
Contents-14
14
ECG and ECG/Respiration Module
Section
This chapter describes the ECG module and the ECG/RESP module. You
will find information on ECG and respiratory monitoring, troubleshooting,
and ST Segment monitoring. It includes the following sections:
• Introduction to ECG Parameter Module. . . . . . . . . . . . . . . . . . . . 14-2
• Considerations When Monitoring ECG . . . . . . . . . . . . . . . . . . . . . 14-7
• ECG Monitoring Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-9
• Respiratory Monitoring Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-18
• ECG Adjustments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-21
• Respiration Adjustments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-39
• ECG and RESP Alarm and INOP Messages . . . . . . . . . . . . . . . . 14-41
• Parameter Settings Transfer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-46
• Accessories and Ordering Information . . . . . . . . . . . . . . . . . . . . 14-47
• ECG/RESP Performance Specifications . . . . . . . . . . . . . . . . . . . 14-50
• Care and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-57
• ST Segment Monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-58
• ST Alarm and INOP Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-67
• Parameter Settings Transfer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-70
• Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-71
Note—
EASI™ is a trademark of Zymed Inc.
ECG and ECG/Respiration Module Section 14-1
Introduction to ECG Parameter Module
Introduction to ECG Parameter Module
ECG and ECG/
Respiration Module
What does it
Measure?
How the
ECG/RESP
Parameter
Works
Monitoring the ECG produc es a continuous waveform of the patient’s
cardiac electrical activity to enable an accurate assessment of his
current physical state.
• The system displays the average Heart Rate (HR), this can be
derived from the ECG or a remote arrhythmia computer.
• You can select HR or PULSE (PULSE can only be selected if you
are monitoring invasive pressure or pleth) as the alarm parameter both share the same alarm limits.
The EASI™ option all o ws yo u to de ri ve a ll 12 standard ECG leads, us in g
a 5-electrode ECG cable and a special electrode placement. This is
described on “EASI™ 12-lead ECG Display, Printing and Recording” on
page 14-33
With the ECG or the ECG/RESP modules listed below, you can use a 5or a 3-electrode ECG cable set to derive up to 8 selectable ECG leads
which can be displayed in up to 3 channels.
Note—
EASI™: With any of the ECG modules listed below and the
EASI™ option switched on, you can use a 5-electrode ECG cable set to
derive the 12 standard ECG leads of which up to 3 can be displayed on
the monitor.
Modules
– M1001A/B ECG Module
– M1002A/B ECG/RESP Module
– CMS Only: M1002A Option C01 ECG/RESP Module with
Analog Respiration Output
• If you are monitoring with a 3-lead set you can select one lead out
of I, II or III to be displayed in 1 channel.
• If you are monitoring with a 5-lead set you can select any two
leads from I, II, III, aVR, aVL, aVF, V or MCL
14-2 ECG and ECG/Respiration Module Section
to be displayed in
1