3B Scientific Rotary-Vane Vacuum Pump User Manual

0 (0)
3B SCIENTIFIC® PHYSICS
Rotary-Vane Vacuum Pump, Two-Stage 1003317
Instruction sheet
1 Oil sight glass 2 Oil drain cap 3 Exhaust filter 4 Oil fill cap 5 Hose nipple 6 Stopcock 7 Ventilation valve 8 Motor 9 Fan cover 10 Power switch and voltage
selector switch 11 Carrying handle 12 Manometer
1. Safety instructions
The rotary-vane pump 1003317 is in accordance with the following applicable European directives and standards: 89/336/EEC Electromagnetic Compatibility 73/23/EEC Low Voltage Directive EMC: EN55014-1 (2000) + A2 (2002), EN55014-2 (1997) + A1 (2001) LVD: EN60034-1 (1998) + A11 (2002)
Safe operation of the pump is guaranteed, pro­vided it is used correctly. However, there is no guarantee of safety if the equipment is used in an improper or careless manner.
The pump may only be used for its intended use. Any other use is to be considered improper.
The manufacturer cannot be held responsible for damages caused by improper use of the pump.
Under no circumstance must the pump be al­tered or tampered with.
In case of tempering the manufacturer declines any responsibility for the functioning and safety of the pump.
If it may be assumed for any reason that non­hazardous operation will not be possible (e.g. visible damage), the pump should be switched off immedi­ately and secured against any unintended use.
In schools and other educational institutions, the operation of the pump must be supervised by qualified personnel.
Before setting up and using the pump for the
first time, read the manual carefully.
Confirm that the voltage selector switch is
set to the local mains voltage.
The pump may only be connected to the mains
via a socket that has an earth connection.
The pump may only be opened/repaired by
qualified and trained personnel. Always discon­nect the pump by pulling the mains plug before proceeding to any of the mentioned operations.
1
In case of power failure during operation,
turn the pump off for safety reasons.
2. Description
The rotary-vane pump 1003317 must only be used in vacuum experiments to depressurize vacuum containers. It is not designed for com­mercial use.
The pump is a high performance, compact, two­stage, oil-sealed rotary vane pump. It is pro­tected against thermal overload and equipped with an exhaust filter to reduce oil mist, a ma­nometer and a
ventilation valve.
For reasons of transport the pump is delivered with no oil in the reservoir. In the package you should find a bottle of oil, sufficient for the first filling.
The pump is equipped with a voltage selector switch (refer to fig. 1), so that it can be operated with mains voltages of 110 V or 230 V plus or minus 10 %.
3. Technical data
Power supply: 110 – 120 V, 60 Hz 220 – 240 V, 50/60 Hz
Suction capacity: 100 l/min Final pressure: 0.003 hPa Motor power: 245 W Oil capacity: 350 ml Manometer: 0 – 1000 mbar Hose nipple: 10 mm dia. Di mensions: approx. 335x138x250 mm
3
Weight: approx. 11 kg Ambient temperature: 5° – 40° C approx.
4. Operation
4.1 General information
After unpacking, please place all parts of the
package (bags, boxes, polystyrene sides) away from the reach of children.
Check the presence of the data label on the
hosing. If the label is not present, do not use the pump and inform the supplier.
In case you need to send back the pump to
the distributer (e.g. repair) drain the oil.
For the disposal of the oil adhere to the local
regulations.
4.2 Before operating the pump
Place the pump horizontally on a stable
support.
Do not connect the pump yet to the mains
voltage.
Check that the voltage selector switch is set
to the local mains voltage and ensure that the power switch is in the OFF position be­fore connecting the pump to a power source.
Remove the oil fill cap and add oil until it
shows at the bottom of the sight glass. Refer to the technical data for the correct oil ca­pacity of the pump.
Screw back the oil fill cap, remove the cap
from the hose nipple and open the stopcock.
Turn the motor switch to ON position.
When the pump runs smoothly put the cap
back onto the hose nipple and close the stopcock. This may take 2 to 30 seconds depending on the ambient temperature.
After the pump operates for approximately
one minute, check the sight glass for proper oil level, which should be aligned with the sight glass oil level line. Refill oil if neces­sary.
Note: Insufficient oil filling will result in poor vac­uum performance. Excessive oil can result in overflowing of oil from the exhaust fitting.
Fig. 1 A voltage selector switch, B power switch, C
socket
4.3 To shut off the pump after use
To prolong life span of the pump and a smooth start-up, these procedures to shut off the pump should be followed.
Close the stopcock.
Remove the hose from the hose nipple.
Switch off the pump and open the stopcock for
a few seconds to relieve any vacuum inside the pump.
Place the cap on the hose nipple to prevent
any contamination or foreign particles from en­tering it.
2