Compressor Nebulizer

Model NE-C28P

Instruction Manual

IM-NE-C28P-E-EN-03-01/2013 1675597-9A

Intended use

Medical Purpose

This product is intended to be used for inhaling medication for respiratory disorders.

Intended User

Legally certified medical experts, such as doctor, nurse and therapist.

Caregiver or patient under the guidance of qualified medical experts for home treatment.

The user should also be capable of understanding general operation of NE-C28P and the content of instruction manual.

Intended Patients

This product should not be used on patients, who are unconscious or are not breathing spontaneously.

Environment

This product is intended for use in a medical facility, such as Hospital, clinic and doctor’s office, and in a room of general household.

Durable period

Durable periods are as follows, provided the product is used to nebulize saline 3 times a day for 10 minutes each time at room

temperature (23°C).

Durable period may vary depending on usage environment.

Compressor (Main unit)

5 years

Nebulizer Kit

1 year

Mouthpiece

1 year

Nosepiece

1 year

Air Tube (PVC, 200cm)

1 year

Air Filter

60 days

Mask (PVC)/(SEBS)

1 year

Precautions for use

Warnings and cautions described in the instruction manual should be observed.

Thank you for purchasing the OMRON COMP AIR.

This product was developed in conjunction with respiratory therapists for the successful treatment of asthma, chronic bronchitis, allergies and other respiratory disorders.

This is a medical device. Operate the device only as instructed by

your doctor and/or respiratory therapist.

How the nebulizer kit works

The medication that is pumped up through the medication channel is mixed with compressed air which is generated by a compressor pump. The compressed air mixed with medication is turned into fine particles and sprayed when in contact with the baffle.

 

Baffle

Aerosol

Aerosol

Nozzle

Medication

Channel

Medication

Medication

 

Compressed Air

 

 

Manufacturer

OMRON HEALTHCARE Co., Ltd.

 

53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN

 

 

EU-representative

OMRON HEALTHCARE EUROPE B.V.

 

Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS

 

www.omron-healthcare.com

 

 

Production facility

OMRON (DALIAN) CO., LTD.

 

Dalian, CHINA

 

 

Subsidiary

OMRON HEALTHCARE UK LTD.

 

Opal Drive, Fox Milne, Milton Keynes, MK15 0DG, U.K.

 

OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH

 

John-Deere-Str. 81a, 68163 Mannheim, GERMANY

 

www.omron-medizintechnik.de

 

OMRON SANTÉ FRANCE SAS

 

14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, FRANCE

 

Made in China

Know your unit

Compressor (Main Unit) x 1

Rear View

 

Nebulizer Kit Holder

 

 

 

 

 

 

 

Ventilation Slots

 

 

Power Cord

Power Plug

 

 

 

 

Air Filter Cover

Ventilation Slots

 

 

 

 

 

 

 

Air Tube Connector

 

 

 

 

Power Switch

 

 

 

 

Nebulizer Kit x 1

Mouthpiece x 1

 

Nosepiece x 1

Cap

 

 

 

 

Inhalation Air Inlet

 

 

 

 

Inhalation Top

 

 

 

 

 

Adult Mask (PVC) x 1

Child Mask (PVC) x 1

Vaporiser Head

Nozzle

Medication Tank

Air Tube Connector

Air Tube

Spare Air Filters x 5

Carrying Bag x 1

(PVC, 200cm) x 1

 

Instruction Manual

 

 

 

 

Warranty card

Important safety instructions

Read all the information in the instruction manual and any other literature included in the box before using the device.

Warning:

Indicates a potentially hazardous situation which, if not avoided, could result in

serious injury.

(Usage)

For type, dose, and regime of medication follow the instructions of your doctor or respiratory therapist.

If you feel anything unusual during use, stop using the device immediately and consult your doctor.

Do not use only water in the nebulizer for inhaling purposes.

Keep the device out of the reach of unsupervised infants and children. The device may contain small pieces that can be swallowed.

Do not store the air tube while there is moisture or medication remaining inside.

Do not use or store the device where it may be exposed to noxious fumes or volatile substances.

Do not use the device where it may be exposed to flammable gas.

Do not cover the compressor with a blanket, towel, or any other type of cover during use.

Always dispose of any remaining medication after use, and use fresh medication each time.

Do not use in anaesthetic or ventilator breathing circuits.

Do not block the ventilation slots. Never place the device where the ventilation slots may be obstructed during operation.

Store the parts in a clean location for avoiding infection.

(Risk of electrical shock)

Never plug or unplug the device with wet hands.

The compressor and power cord are not waterproof. Do not spill water or other liquids on these parts. If liquid does spill on these parts, immediately unplug the power cord and wipe off the liquid with gauze or other absorbent material.

Do not immerse the compressor in water or other liquid.

Do not use or store the device in humid locations.

Do not operate the device with a damaged power cord or plug.

Keep the power cord away from heated surfaces.

(Cleaning and disinfecting)

Observe the rules below when cleaning or disinfecting parts. Failure to observe these rules may result in damage, inefficient nebulization or infection. For the instructions, refer to “Cleaning and disinfecting” section.

Clean and disinfect the nebulizer kit, mask, mouthpiece or nosepiece before using them when as follows:

-the first time after purchase

-if the device has not been used for a long period of time

-if more than one person uses the same device

Be sure to wash or wipe the parts after use, and ensure that they are thoroughly disinfected and dried, and stored in a clean location.

Do not leave the cleaning solution in the parts.

Caution:

Indicates a potentially hazardous situation which if not avoided, may result in

minor or moderate injury, or physical damage.

(Usage)

Provide close supervision when this device is used by, on, or near children or invalids.

Make sure that the parts are attached correctly.

Make sure that the air filter is correctly attached.

Make sure that the air filter is clean. If the air filter has been used for more than 60 days, replace it with a new one.

Do not tilt the nebulizer kit at an angle of greater than 45 degrees in all directions or shake it while in use.

Do not use or store the device while the air tube is creased.

Use only original nebulizing parts, air tube, air filter and air filter cover.

Do not add more than 7 ml of medication to the medication tank.

Do not carry or leave the nebulizer kit while the medication tank contains medication.

Do not leave the device unattended with infants or persons who cannot express their consent.

Do not subject the device or the parts to any strong shocks such as dropping the device on the floor.

Do not distort the vaporiser head, or jab the nozzle of the medication tank with a pin or any sharp object.

Do not insert fingers or objects inside the compressor.

Do not disassemble or attempt to repair the compressor or power cord.

Do not leave the device or its parts in extreme hot or cold temperature, or under direct sunlight.

Do not block the air filter cover.

Do not use the device while sleeping or if drowsy.

Approved for human use only.

The compressor may become hot during operation.

Do not touch the compressor for other than necessary operation such as turning off the power while nebulizing.

To avoid the medication residue on the face, be sure to wipe the face after removing the mask.

To avoid injury to the nose mucosa, do not squeeze the nosepiece into the back of the nose.

Do not block the slit between the cap and the inhalation air inlet.

Make sure that the vaporiser head is correctly installed before use.

Do not use a damaged nebulizer kit, mouthpiece or nosepiece.

(Risk of electrical shock)

Always unplug the device from the power outlet after use and before cleaning.

Plug the device into the appropriate voltage outlet. Do not overload power outlets or use extension cords.

Do not misuse the power cord.

Do not wind the power cord around the compressor.

Changes or modifications not approved by OMRON HEALTHCARE will void the user warranty.

(Cleaning and disinfecting)

Observe the rules below when cleaning or disinfecting parts. Failure to observe these rules may result in damage, inefficient nebulization or infection. For the instructions, refer to “Cleaning and disinfecting” section.

Do not use a microwave oven, dish dryer or hair dryer to dry the device or the parts.

Do not use an autoclave, EOG gas sterilizer or low temperature plasma sterilizer.

When disinfecting parts by boiling, make sure that the container does not boil dry. Otherwise it may also result in fire.

General Safety Precautions:

Inspect the device and parts before using them each time, and check that there are no problems. In particular, be sure to check the following:

-That the nozzle or air tube are not damaged.

-That the nozzle is not blocked.

-That the compressor operates normally.

When using this device, there will be some noise and vibration caused by the pump in the compressor. There will also be some noise caused by the emission of compressed air from the nebulizer kit. This is normal and does not indicate a malfunction.

Operate the device only as intended. Do not use the device for any other purpose.

Do not use the device at temperatures greater than +40°C.

Make sure that the air tube is securely attached to the compressor and nebulizer kit, and does not come loose. Twist the air tube plug slightly when inserting it into the air tube connectors to avoid the air tube disconnecting during use.

To completely isolate the device from the power source unplug the plug from the power source.

Keep these instructions for future reference.

Important information regarding Electro Magnetic Compatibility (EMC)

With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation.

Medical devices should also not interfere with other devices.

In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe product situations, the EN60601-1-2:2007 standard has been implemented. This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices.

This medical device manufactured by OMRON HEALTHCARE conforms to this EN60601-1-2:2007 standard for both immunity and emissions.

Nevertheless, special precautions need to be observed:

Do not use mobile (cellular) telephones and other devices, which generate strong electrical or electromagnetic fields, near the medical device. This may result in incorrect operation of the unit and create a potentially unsafe situation. Recommendation is to keep a minimum distance of 7 m. Verify correct operation of the device in case the distance is shorter.

Further documentation in accordance with EN60601-1-2:2007 is available at OMRON HEALTHCARE EUROPE at the address mentioned in this instruction manual.

Documentation is also available at www.omron-healthcare.com.

Correct Disposal of This Product

(Waste Electrical & Electronic Equipment)

This marking shown on the product or its literature, indicates that it should not be

disposed of, with other household wastes at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this product from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources.

Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can return this item for environmentally safe recycling.

Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial wastes for disposal.

Optional Medical Accessories

(within the scope of EC Medical Device Directive 93/42/EEC)

Name

Model

Order No.

V.V.T. Nebulizer Kit N-01 &

NEB-NSET3-81E

9515573-6

Mouthpiece Set

 

 

 

 

 

Adult Mask (PVC)

NEB-MSLP-E

9956275-1

 

 

 

Child Mask (PVC)

NEB-MSMP-E

9956276-0

 

 

 

Adult Mask (SEBS)

NEB-MSLS-E

9956312-0

 

 

 

Child Mask (SEBS) Set

NEB-MSSS-E

9956281-6

 

 

 

Mouthpiece

NEB-MP-E

9956273-5

 

 

 

Nosepiece

NEB-NP-E

9956274-3

 

 

 

Air Tube (PVC, 200cm)

NEB-TP-L20E

9520606-3

 

 

 

Air Tube (Silicone, 100cm)

NEB-TSC-L10E

9956269-7

 

 

 

Other Optional/Replacement Parts

Name

Model

Order No.

Air Filters x 5

NEB-AFR-28E

9956271-9

 

 

 