Omron MC-514 User Manual

0 (0)
INSTRUCTION MANUAL
MC-514
EAR THERMOMETER
Model MC-514
ESPAÑOL
ENGLISH
TABLE OF CONTENTS
Introduction ................................................................................ 3
Part Names ................................................................................ 4
Conversion °F/°C ...................................................................... 4
Before Use ................................................................................ 5
Directions for Use and for Memory .......................................... 5
Care and Maintenance ................................................................ 6
Battery Replacement .................................................................. 6
Specifications ............................................................................ 7 Troubleshooting
Warning .................................................................................... 10
FCC Statement ........................................................................ 11
Warranty .................................................................................. 12
Notes .................................................................................. 13-15
Quick Reference ...................................................................... 16
............................................................................8,
9
INTRODUCTION
Thank you for purchasing the Omron Ear Thermometer.
Fill-in for future reference
DATE PURCHASED: __________________ SERIAL NUMBER: ____________________
Staple your purchase receipt here.
This ear thermometer offers a safe, accurate and quick temperature reading from the ear canal. This thermometer senses the infrared heat from the ear drum (tympanic membrane).
No need to worry about broken glass or mercury hazards. An ear temperature is not affected by talking, eating or drinking.
These factors do affect an oral temperature. In one second it changes the heat measured into an equivalent temperature. The unit is easy to use for children and adults. Temperatures can
be taken when a child is sleeping. We recommend you practice taking a temperature with the Omron
Ear Thermometer on each member of your family when they are healthy. This way you will be confident in your technique when using the ear thermometer on a sick family member.
Contact your doctor if you have any questions regarding specific temperatures.
This product is intended for home use only.
Read this instruction manual completely before using your
Omron Ear Thermometer.
SAVE THESE INSTRUCTIONS
PART NAMES
MC-514
TITLE
Lens
Lens cover
Probe
Storage case (with 10 lens covers inside, plus 1 on the thermometer)
Lens
Probe
Box for lens covers (with 10 lens covers inside)
Button
Display
CONVERSION BETWEEN
FAHRENHEIT to CENTIGRADE (°F/°C)
• The thermometer is initially set at Fahrenheit °F
• Follow the steps below to change the measurement scale from °F to Centigrade °C:
Note:
Changing the measurement scale the previous measured valued.
1) Be sure unit is in the off position (blank display).
2) Press and hold the Button for 3 seconds until the "°F"
flashes and you hear one beep.
3) Press button again, now "°C" flashes.
4) After five (5) seconds the unit will shut off.
• The desired mode has been set.
• Repeat the process to reverse measurement scale if desired.
“°F” to “°C” or “°C” to “°F” deletes
TITLE
BEFORE USE
• The thermometer will not work unless a lens cover is applied correctly.
• For hygienic use, attach a new lens cover each time. REPLACING LENS COVER
Use Omron lens cover (model MC-505LC).
1) Remove the lens cover.
Insert the lens cover until it clicks.
2) Attach a new lens cover.
DIRECTIONS FOR USE AND MEMORY
(1) Push button to turn on. (The display will illuminate for 2 seconds.)
Memory Feature:
(2) The previous value measured is shown on the display (Memory). For the 2nd
and following measurements, the previous value measured is shown with “M” symbol (Memory).
Note: When the unit is used for the first time the “M” symbol is not displayed.
(3) When “°F” or “°C” symbol flashes - insert lens into ear. Gently pull back the ear
lobe before inserting for best results. Tugging on the earlobe provides a direct line to the eardrum to maximize the accuracy of results.
(4) Listen for ready beep (1 beep); an “O” symbol will appear indicating its reading.
Angle is correct; ready for measurement.
(5) Press button again while unit is placed in the ear, 4 beeps means
measurement is complete.
(6) Remove probe from ear. (7) Discard used lens cover, and replace with a new one,
so that unit is ready for use again.
(8) If you are going to repeat measurements, wait until “°F” or “°C” flashes.
Note:
- You can measure consecutively up to three times. If you are going to measure more than this, wait ten minutes, then measure again. This is because the measurement may be inaccurate due to the probe warmed by body heat.
- The thermometer will automatically turn off about 1 minute after a temperature has been taken. There is no button to turn the power off manually.
CARE AND MAINTENANCE
• How to Clean the Unit
- Wipe off the casing of the unit with a soft, dry cloth, or with rubbing alcohol
- When dirty clean the lens, only use a soft dry
cloth or cotton swab.
- Do not clean any part of the unit with
detergent, thinner or benzene.
- Do not immerse in water.
• How to Store your thermometer
- Keep your thermometer in the storage case after attaching a lens cover to protect the lens.
- Store your thermometer in a place that is out of direct sunlight, is not in any extremes of temperature or humidity and is not likely to be affected by vibration, dust or shock.
BATTERY REPLACEMENT
(1) Use a lithium battery CR2032. (2) Remove the screw and battery cover. (3) Remove old battery. Do not use a screw driver or metal
tweezers to remove battery. (4) Place the new battery with the positive (+) side up. (5) Replace the battery cover and tighten the screw. Note: Dispose of any used battery in
accordance with local regulations.
TITLE
(4)
(2)
(5)
(3)
SPECIFICATIONS
Name: Omron Ear Thermometer Model: MC-514 Power Source: DC 3 Volt (lithium battery CR2032)
Battery life: Approximately 2000 measurements Sensor: Thermopile
Temperature indication: In 4 digits, every 0.1°F or °C Accuracy: 93.2 ~ 96.7°F
Measurement range Weight: Approx. 1.8 oz (50 g) (including the battery) Operating Ambient temperature; 50 to 104°F
temperature / humidity: Relative humidity; 30 to 85% Storage Ambient temperature; -4 to 140°F
temperature / humidity: Relative humidity; 10 to 95% Memory: Memory function is to display the previous reading Contents: Thermometer, Storage Case, Instruction Manual,
Options: Use Omron model MC-505LC
The specifications may be changed without prior notice.
: 93.2 to 108°F (34 to 42.2°C)
TITLE
(34.0 ~ 35.9°C) : ±0.5°F (±0.3°C)
96.8 ~ 102.2°F (36.0 ~ 39.0°C) : ±0.4°F(±0.2°C)
102.3 ~ 108.0°F (39.1 ~ 42.2°C) : ±0.5°F(±0.3°C)
(10 to 40°C)
(-20 to 60°C)
and 21 Lens Covers
TROUBLESHOOTING
TITLE
Display
Nothing is displayed even after unit is turned on.
“°F” or “°C”flashes, but “0” symbol is NOT displayed
A "0 0" symbol is displayed.
After all the symbols on display illuminate, a " " mark is displayed.
Meaning Action
Dead battery. Replace battery.
Polarities of battery (+ and -) are wrong.
The lens cover may be dirty.
Lens may be dirty.
Ear drum may be blocked with earwax.
The unit is working under high room temperature, where a special operation is needed.
Reposition the battery with correct polarities.
Replace the lens cover with a new one.
Clean the lens with a cotton swab.
Remove earwax.
Place unit in ear, wait 2 seconds. Press button and wait for 4 beeps. Remove from ear, look at results.
Replace battery.Low battery.
The display light is not turned on.
Low battery. Replace battery.
TROUBLESHOOTING
TITLE
Meaning ActionDisplay
Lens cover is not on Probe.
Attach the lens cover, then measure.
Temperature of the unit is unstable.
The temperature of the unit is
50°F
104°F
or below
or above
which is outside the operating range.
The temperature of the unit is which is outside the operating range.
May require service.
The measured temperature was above the range of the thermometer.
The measured temperature was below the range of the thermometer.
Make sure unit remains at room temperature where measurement will take place. Allow 30 minutes to stabilize before using.
Normal range: (50°F to 104°F) (10°C to 40°C)
Turn on the unit again. If "Er.1" is displayed again, the unit may require service. Call Customer Service Dept. 1-800-634-4350
Probe may not have been properly positioned in ear. The probe must be snug against opening of ear. Be sure lens cover is clean and ear canal is not blocked with earwax.
WARNING
TITLE
• This thermometer is intended to be used in human ears only.
• Only use the lens cover for this specific unit.
• For proper hygiene DO NOT
• Replace the lens cover when it is effected by ear wax or other body fluids.
• Handle the lens with care. Improper handling may effect
temperature accuracy.
• Keep the unit out of the reach of children.
• Use your unit in room temperature. Extremes in environmental
temperature may effect the accuracy.
• Store unit at room temperature when possible. Keep thermometer
at room temperature for at least 30 minutes before use.
• Remove unit if pain is noted during use.
• DO NOT
• DO NOT
tissue as well as effect the accuracy of the temperature.
• DO NOT
• DO NOT
• DO NOT
use if ear infection is present. use right after swimming or bathing as this may injure ear
use near a portable phone. take apart, repair or change any parts of the unit at any time.
get the unit wet or pour any liquid onto any portion of the unit.
share lens covers.
1010