OMEGA
HH508/HH509/HH509R
Digital Thermometer
omega.comTM
|
|
OMEGAnetSM On-Line Service |
Internet e-mail |
|
|
||
|
|
|
http://www.omega.com |
info@omega.com |
|
|
|
|
|
|
Servicing North America: |
||||
USA: ISO 9001 Certified |
Canada: |
||||||
|
|
|
|
|
|
||
One Omega Drive, Box 4047 |
976 Bergar |
||||||
Stamford, CT 06907-0047 |
Laval (Quebec) H7L5A1 |
||||||
Tel: (203) 359-1660 |
Tel: (514) 856-6928 |
||||||
FAX: (203)359-7700 |
FAX: (514) 856-6886 |
||||||
e-mail: info@omega.com |
e-mail: info@omega.com |
||||||
|
|
For immediate technical or application assistance: |
|||||
USA and Canada: |
Mexico and Latin America: |
||||||
Sales Service: 1-800-826-6342 / 1-800-TC-OMEGASM |
Tel: (95) 800-TC-OMEGASM |
||||||
Customer Service: 1-800-622-2378 / 1-800-622-BESTSM |
FAX: (95) 203-359-7807 |
||||||
Engineering Service: 1-800-872-9436 / 1-800-USA-WHENSM |
En Español: (203) 359-7803 |
||||||
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA |
e-mail: espanol@omega.com |
||||||
|
|
|
Servicing Europe: |
||||
Benelux: |
Germany/Austria: |
||||||
Postbus 8034, 1180 LA Amstelveen, |
Daimlerstrasse 26, D-75392 |
||||||
The Netherlands |
Deckenpfronn, Germany |
||||||
Tel: (31) 20 6418405 FAX: (31) 20 6434643 |
Tel: 49 (07056) 3017 FAX: 49 (07056) 8540 |
||||||
Toll Free in Benelux: 06 0993344 |
Toll Free in Germany: 0800 82 66342 |
||||||
e-mail: nl@omega.com |
e-mail: germany@omega.com |
||||||
Czech Republic: |
United Kingdom: ISO 9002 Certified |
||||||
ul. Rude armady 1868, 733 01 Karvina- |
One Omega Drive |
|
|
||||
Hranice, Czech Repubic |
Riverbend Technology Centre Northbank, Irlam, |
||||||
Tel: 420 (69) 6311627 FAX: 420 (69) 6311114 |
Manchester, M44 5EX, England |
||||||
e-mail: czech@omega.com |
Tel: 44 (161) 777-6611 FAX: 44 (161) 777-6622 |
||||||
France: |
|
|
|
|
|||
9, rue Denis Papin, 78190 Trappes |
Toll Free in England: 0800-488-488 |
||||||
Tel: (33) 130-621-400 FAX: (33)130-699-120 |
e-mail: sales@omega.com.uk |
||||||
Toll Free in France: 0800-4-06342 |
|
|
|
|
|||
e-mail: france@omega.com |
|
|
|
|
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, patient connected application.
INTRODUCTION
This instrument is a 3½ digit, compact-sized portable digital thermometer designed to use external K/J/T/E/S type thermocouples as temperature sensor. Temperature indication follows Reference Temperature/Voltage Tables (N.I.S.T. Monograph 175 Revised to ITS-90) for K/J/T/E/S type thermocouples.
SAFETY INFORMATION
It is recommended that you read the safety and operation instructions before using the thermometer.
WARNING
To avoid electrical shock, do not use this instrument when working voltages at the measurement surface over 24V AC or DC.
WARNING
To avoid damage or burns, do not make temperature measurement in microwave ovens.
CAUTION
Repeated sharp flexing can break the thermocouple leads. To prolong lead life, avoid sharp bends in the leads, especially near the connector.
The symbol on the instrument indicates that the operator must refer to an explanation in this manual.
1
SPECIFICATIONS
ENVIRONMENTAL
Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H. Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H.
GENERAL
Display: 3½ digit liquid crystal display (LCD) with a maximum reading of 1999. Polarity: Automatic, positive implied, negative polarity indication. Overrange: -OL is displayed.
Zero: Automatic.
Reading Rate: one time per second.
Low battery indication: the " " is displayed when the battery voltage drops below the operating level.
Measurement rate: 1 times/second.
Accuracy: Stated accuracy at 23°C±5°C, <75% relative humidity. Dimensions: 195mm(H) x 92mm(W) x 53mm(D).
Weight: 365g.
Input Connector: Accepts standard miniature thermocouple connectors (flat blades spaced 7.9mm, center to center).
Battery Life: 100 hours typical battery.
Auto power off: The meter key switch inactive for more than 30 minutes, press power key to resume operation.
Temperature Coefficient: 0.1 times the applicable accuracy specification per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F).
Input Protection: 24V dc or 24V ac rms maximum input voltage on any combination of input pins.
Maximum Differential Common Mode Voltage (Maximum Voltage between T1 and T2 during measurement): 1volt.
2
ELECTRICAL
Temperature Scale: Celsius or Fahrenheit user-selectable.
Measurement Range:
Thermocouple |
Range |
K-TYPE(0.1°C) |
-200°C to 1372°C, -328°F to 1999°F |
J-TYPE(0.1°C) |
-210°C to 1200°C, -346°F to 1999°F |
T-TYPE(0.1°C) |
-200°C to 400°C, -328°F to 752°F |
E-TYPE(0.1°C) |
-220°C to 1000°C, -364°F to 1832°F |
S-TYPE(1°C) |
0°C to 1767°C, 32°F to 1999°F |
Auto range: 0.1°C/1°C, 0.1°F/1°F.
Accuracy: Accuracy is specified for operating temperatures over the range of 18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error.
|
±(0.1% rdg + 1°C) range -60°C to 1372°C |
|
±(0.1% rdg + 2°C) range -60°C to -220°C |
|
±(0.1% rdg + 2°F) range -76°F to 1999°F |
|
±(0.1% rdg + 4°F) range -76°F to -364°F |
S-TYPE |
±(0.0% rdg + 2°C) on 0°C to 50°C |
|
±(0.1% rdg + 1°C) on 50°C to 1767°C |
|
±(0.0% rdg + 4°F) on 32°F to 122°F |
|
±(0.1% rdg + 2°F) on 122°F to 1999°F |
3
OPERATING INSTRUCTIONS
1 Power Switch
The key turns the thermometer on or off. In the MIN MAX record mode can not power off, must leave MIN MAX record mode then power off.
25 °C °F Selecting the Temperature Scale
Readings is displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F). When the thermometer is turned on, it is set to the temperature scale that was in use when the thermometer was last turned off. To change the temperature scale, press the °C or °F key.
3"" Button
Press "" button to toggle on and off of backlight. The backlight will switchoff automatically after 30 seconds.
4HOLD Mode
Pressing the HOLD key to enter the Data Hold mode, the "HOLD" annunciator is displayed. When HOLD mode is selected, the thermometer held the present readings and stops all further measurements.
Pressing the HOLD key again to cancel HOLD mode causing thermometer to resume taking measurements.
In the MIN/MAX recording mode, press HOLD key to stop the recording. Press HOLD key again to resume recording. (Previously recorded reading are not erased).
4