HP 8753ET, 8753ES User Manual

0 (0)

specifications

8753ET/ES

Network Analyzers

HP 8753ET and 8753ES Network Analyzers 30 kHz to 3 or 6 GHz

This document describes the performance and features of the following products:

HP 8753ES S-parameter vector network analyzer

HP 8753ES Option 011 vector network analyzer without test set

HP 8753ET Transmission/reflection vector network analyzer

For more information about these analyzers, please see the following documents:

 

HP literature number

HP 8753ET and 8753ES Network Analyzers Brochure

5968-5159E

HP 8753ET and 8753ES Network Analyzers Configuration Guide

5968-5158E

Table of contents

Definitions and test conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 System performance summaries . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 Specifications and characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . .10 Measurement throughput summary . . . . . . . . . . . . . . . . . . . . . . . . .21 Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21 Test set specifications for Option 011 . . . . . . . . . . . . . . . . . . . . . . . .22 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24

2

Definitions and test conditions

Specifications describe the instrument's warranted performance after a half-hour warm-up and over the temperature range of 25 ˚C ± 5 ˚C, unless otherwise stated. Specifications for frequencies above 3 GHz do not apply to instruments with Option 075 (75-ohm impedance).

Supplemental characteristics are typical but non-war- ranted performance parameters. These are denoted as “typical,” “nominal,” or “approximate.”

The measurement uncertainty curves and measurement port characteristics given for HP 8753ES systems also apply to the HP 8753ES with Options 006 and 011 and the HP 85047A test set (50-ohm), or the HP 8753ES Option 011 with an HP 85046B test set (75-ohm).

Dynamic range

System dynamic range is calculated as the difference between the receiver noise floor and the lesser of either the source maximum output or the receiver maximum input level. System dynamic range applies to transmission measurements only, since reflection measurements are limited by directivity.

Noise floor is specified as the mean of the noise trace over frequency. Noise floor is measured with the test ports terminated in loads, full two-port error correction for the HP 8753ES and enhanced-response error correction for the HP 8753ET (with 16 averages used during isolation), 10 Hz IF bandwidth (BW), maximum test port power, and no averaging during the measurement.

Measurement uncertainty

Measurement uncertainty curves utilize a Root Sum Square (RSS) model for the contribution of random errors such as noise, typical connector repeatabilities, and test set switching; this is combined with a worstcase model for the contributions of dynamic accuracy and residual systematic errors.

Curves show the worst-case magnitude and phase uncertainty for reflection and transmission measurements, after a full two-port error correction for the HP 8753ES and enhanced-response error correction for the HP 8753ET (including isolation with an averaging factor of 16) using the specified cal kit, with 10 Hz IF bandwidth (BW) and no averaging.

Measurement port characteristics

Characteristics show the residual system uncertainties for uncorrected performance and after accuracy enhancement using full two-port error correction for the HP 8753ES and enhanced-response error correction for the HP 8753ET. These characteristics apply for an environmental temperature of 25 ± 5 ˚C, with less than 1 ˚C deviation from the calibration temperature. HP 8753ET Option 004 may degrade transmission source match as much as 2 dB, resulting in up to 0.05 dB additional uncertainty in transmission tracking.

Corrected performance indicates residual error after calibration. It is determined by the quality of calibration standards, system repeatability, stability, and noise.

Uncorrected performance indicates intrinsic errors without calibration correction applied. This is related to the ultimate stability of a calibration.

3

System performance summaries

HP 8753ES (50-ohm systems)

7-mm test ports

Configuration

 

Network analyzer

HP 8753ES

 

Standard and Option 006

Calibration kit

HP 85031B

Test-port cables

HP 11857D

System dynamic range

 

30 kHz to 50 kHz

70 dB1

50 kHz to 300 kHz

90 dB1

300 kHz to 16 MHz

100 dB

16 MHz to 3 GHz

110 dB

Option 014

108 dB

3 GHz to 6 GHz

105 dB

Option 014

103 dB

Reflection uncertainty

Magnitude

Transmission uncertainty

Magnitude

1.Typical below 300 kHz.

2.Typical performance.

Measurement port characteristics

Frequency range

Corrected 30 kHz-300 kHz2

300 kHz-1.3 GHz

1.3 GHz-3 GHz

3 GHz- 6 GHz

Directivity

55 dB

55 dB

51 dB

46 dB

Source match

55 dB

51 dB

49 dB

43 dB

Load match

55 dB

55 dB

51 dB

46 dB

Reflection

±(0.001 dB

±(0.001 dB

±(0.005 dB

±(0.020 dB

tracking

+0.02 dB/˚C)

+0.01 dB/˚C)

+0.02 dB/˚C)

+0.03 dB/˚C)

Transmission

±(0.008 dB

±(0.006 dB

±(0.009 dB

±(0.021 dB

tracking

+0.02 dB/˚C)

+0.01 dB/˚C)

+0.02 dB/˚C)

+0.03 dB/˚C)

Uncorrected

 

 

 

 

Directivity

20 dB*

35 dB

30 dB

25 dB

Source match

18 dB**

16 dB

16 dB

14 dB

Load match

18 dB**

18 dB

16 dB

14 dB

Reflection

±2.5 dB

±1 dB

±1 dB

±1.5 dB

tracking

 

 

 

 

Transmission

±2.5 dB

±1 dB

±1 dB

±1.5 dB

tracking

 

 

 

 

Crosstalk

90 dB***

100 dB

100 dB

90 dB

*15 dB, 30 kHz to 50 kHz

**10 dB, 30 kHz to 50 kHz

 

***60dB, 30 kHz to 50 kHz

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Phase

Phase

4

System performance summaries

HP 8753ES (50-ohm systems)

Type-N test ports

Configuration

 

Network analyzer

HP 8753ES

 

Standard and Option 006

Calibration kit

HP 85032B

Test-port cables

HP 11857D

System dynamic range

 

30 kHz to 50 kHz

70 dB1

50 kHz to 300 kHz

90 dB1

300 kHz to 16 MHz

100 dB

16 MHz to 3 GHz

110 dB

Option 014

108 dB

3 GHz to 6 GHz

105 dB

Option 014

103 dB

Reflection uncertainty

Magnitude

Transmission uncertainty

Magnitude

1.Typical below 300 kHz.

2.Typical performance.

Measurement port characteristics

Frequency range

Corrected

30 kHz-300 kHz2

300 kHz-1.3 GHz

1.3 GHz-3 GHz

3 GHz-6 GHz

Directivity

 

50 dB

50 dB

47 dB

40 dB

Source match

49 dB

42 dB

36 dB

31 dB

Load match

 

50 dB

50 dB

47 dB

40 dB

Reflection

 

±(0.005 dB

±(0.009 dB

±(0.019 dB

±(0.070 dB

tracking

 

+0.02 dB/˚C)

+0.01 dB/˚C)

+0.02 dB/˚C)

+0.03 dB/˚C)

Transmission

 

±(0.014 dB

±(0.013 dB

±(0.026 dB

±(0.065 dB

tracking

 

+0.02 dB/˚C)

+0.01 dB/˚C)

+0.02 dB/˚C)

+0.03 dB/˚C)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Phase

Phase

5

HP 8753ET, 8753ES User Manual

System performance summaries

HP 8753ES (50-ohm systems)

3.5-mm test ports

Configuration

 

Network analyzer

HP 8753ES

 

Standard and Option 006

Calibration kit

HP 85033D

Test-port cables

HP 11857D

System dynamic range

 

30 kHz to 50 kHz

70 dB1

50 kHz to 300 kHz

90 dB1

300 kHz to 16 MHz

100 dB

16 MHz to 3 GHz

110 dB

Option 014

108 dB

3 GHz to 6 GHz

105 dB

Option 014

103 dB

Reflection uncertainty

Magnitude

Transmission uncertainty

Magnitude

1.Typical below 300 kHz.

2.Typical performance.

Measurement port characteristics

Frequency range

Corrected 30 kHz-300 kHz2 300 kHz-1.3 GHz

1.3 GHz-3 GHz

3 GHz-6 GHz

Directivity

49 dB

46 dB

44 dB

38 dB

Source match

49 dB

44 dB

41 dB

37 dB

Load match

49 dB

46 dB

44 dB

38 dB

Reflection

±(0.010 dB

±(0.005 dB

±(0.007 dB

±(0.009 dB

tracking

+0.02/˚C)

+0.01/˚C)

+0.02/˚C)

+0.03/˚C)

Transmission

±(0.016 dB

±(0.014 dB

±(0.022 dB

±(0.048 dB

tracking

+0.02/˚C)

+0.01/˚C)

+0.02/˚C)

+0.03/˚C)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Phase

Phase

6

System performance summaries

HP 8753ES (75-ohm systems)

Type-N test ports

Configuration

 

Network analyzer

HP 8753ES Option 075

Calibration kit

HP 85036B

Test-port cables

HP 11857B

System dynamic range

 

30 kHz to 50 kHz

68 dB1

50 kHz to 300 kHz

90 dB1

300 kHz to 16 MHz

96 dB

16 MHz to 3 GHz

106 dB

Reflection uncertainty

Magnitude

Transmission uncertainty

Magnitude

1.Typical below 300 kHz.

2.Typical performance.

Measurement port characteristics

Frequency range

Corrected

30 kHz - 300 kHz2

300 kHz - 1.3 GHz

1.3 GHz - 3 GHz

Directivity

 

48 dB

48 dB

43 dB

Source match

 

47 dB

41 dB

35 dB

Load match

 

48 dB

48 dB

43 dB

Reflection

 

±(0.004 dB

±(0.010 dB

±(0.019 dB

tracking

 

+0.02 dB/˚C)

+0.01 dB/˚C)

+0.02 dB/˚C)

Transmission

 

±0.018 dB

±(0.016 dB

±(0.033 dB

tracking

 

+0.02 dB/˚C)

+0.01 dB/˚C)

+0.02 dB/˚C)

Uncorrected2

30 kHz - 300 kHz

300 kHz - 1.3 GHz

1.3 GHz - 3 GHz

Directivity

 

20 dB*

35 dB

30 dB

Source match

 

16 dB**

16 dB

16 dB

Load match

 

15 dB**

18 dB

16 dB

Reflection

 

±2.5 dB

±1 dB

±1 dB

tracking

 

 

 

 

 

Transmission

 

±2.5 dB

±1 dB

±1 dB

tracking

 

 

 

 

 

Crosstalk

 

90 dB***

100 dB

100 dB

* 15 dB, 30 to 50 kHz

**10 dB, 30 to 50 kHz *** 60 dB, 30 to 50 kHz

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Phase

Phase

7

System performance summaries

HP 8753ES (75-ohm systems)

Type-F test ports

Configuration

 

Network analyzer

HP 8753E Option 075

Calibration kit

HP 85039B

Test-port cables

HP 11857B

System dynamic range

 

30 kHz to 50 kHz

68 dB1

50 kHz to 300 kHz

90 dB1

300 kHz to 16 MHz

96 dB

16 MHz to 3 GHz

106 dB

Reflection uncertainty

Magnitude

Transmission uncertainty

Magnitude

1.Typical below 300 kHz.

2.Typical performance.

Measurement port characteristics

Data is shown for a Type-F female reflection port and a Type-F male transmission port.

Frequency range

Corrected2

30 kHz-300 kHz

300 kHz-1.3 GHz

1.3 GHz-3 GHz

Directivity

45 dB

45 dB

40 dB

Source match

40 dB

40 dB

30 dB

Load match

45 dB

45 dB

40 dB

Reflection

±(0.060 dB

±(0.060 dB

±(0.024 dB

tracking

+0.02 dB/˚C)

+0.01 dB/˚C)

+0.02 dB/˚C)

Transmission

±(0.033 dB

±(0.019 dB

±(0.057 dB

tracking

+0.02 dB/˚C)

+0.01 dB/˚C)

+0.02 dB/˚C)

 

 

 

 

 

 

 

 

 

 

 

 

Phase

Phase

8

System performance summaries

HP 8753ET (50-ohm systems)

Type-N test ports

Configuration

 

 

 

 

Network analyzer

HP 8753ET Standard,

 

 

Option 006 or Option 004

Calibration kit

 

HP 85032B

Test port cable

 

HP part number 8120-4781

System dynamic range

 

 

 

300 kHz to 16 MHz

100 dB

 

 

16 MHz to 3 GHz

110 dB

 

 

3 GHz to 6 GHz

 

105 dB

 

 

Measurement port characteristics

 

 

 

 

 

 

Frequency range

Uncorrected 300 kHz-1.3 GHz 1.3 GHz-3.0 GHz

3.0 GHz-6.0 GHz

Directivity

30 dB

24 dB

19 dB

Source match4

25 dB

20 dB

14 dB

Load match

24 dB

19 dB

16 dB

Reflection tracking

±1.0 dB

±1.0 dB

±2.0 dB

Transmission tracking

±1.5 dB

±1.5 dB

±2.5 dB

Crosstalk

100 dB

100 dB

90 dB

 

Transmission uncertainty: (enhanced-response calibration)

Transmission uncertainty: (response calibration)

Reflection uncertainty (one-port calibration)

Frequency range (continued)

 

Corrected

300 kHz-1.3 GHz

1.3 GHz-3.0 GHz 3.0 GHz-6.0 GHz

 

Reflection measurements1

 

 

 

Directivity

 

50 dB

47 dB

40 dB

Source match

 

42 dB

36 dB

31 dB

Load match

 

 

 

 

 

One-port cal

 

24 dB

19 dB

16 dB

 

Enhanced reflection cal

24 dB

19 dB

16 dB

Reflection tracking

 

±(0.009 dB

±(0.019 dB

±(0.07 dB

 

 

 

+0.01 dB/˚C)

+0.02 dB/˚C)

+0.03 dB/˚C)

Transmission measurements2

 

 

Source match

 

 

 

 

 

Enhanced-response cal

42 dB

36 dB

31 dB

 

Response-only cal3

 

25 dB

20 dB

14 dB

Transmission tracking

 

 

 

 

 

Enhanced-response cal

±(0.006 dB

±(0.018 dB

±(0.054 dB

 

 

 

+0.01 dB/˚C)

+0.02 dB/˚C)

+0.03 dB/˚C)

 

Response-only cal

 

±(0.033 dB

±(0.1 dB

±(0.27 dB

 

 

 

+0.01 dB/˚C)

+0.02 dB/˚C)

+0.03 dB/˚C)

1.

One-port or enhanced-response calibration.

 

2.

Enhanced-response or response-only calibration.

 

3.

2 dB less with Option 004.

 

4.

Option 004 may degrade uncorrected source match as much as 2 dB.

 

9

 

 

 

Loading...
+ 19 hidden pages