Hanna Instruments HI8417, HI8519, HI8520, HI8521 User Manual

4 (1)

Instruction Manual

HI8417 - HI8519 HI8520 - HI8521

Microprocessor

Bench-top

pH Meters

HI 8417

°C

pH

OFF

ON

 

 

 

 

 

BUFFER V

 

 

 

 

 

 

 

 

t1

d1

 

NBS

ALUES

20°C

 

 

 

 

d2

 

 

STD

pH

 

6,88

 

 

 

 

 

NBS

 

pH

 

 

 

 

 

t2

 

pH

 

7,02

 

 

 

 

d3

 

STD

 

4,00

 

 

CAL

 

t3

NBS

 

 

pH

 

 

°C

 

pH

 

4,00

 

 

 

 

 

 

STD

 

9,22

 

 

 

 

 

 

 

 

 

pH

 

 

 

 

 

 

 

 

 

 

 

10,05

 

 

CON

 

 

 

 

 

 

 

 

 

 

 

 

 

°C

 

 

 

 

 

 

 

 

 

 

 

 

pH mV

 

 

 

 

 

 

 

 

These Instruments are in

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

h t t p : / / w w w . h a n n a i n s t . c o m

Compliance with the CE Directives

Dear Customer,

Thank you for choosing a Hanna Instruments Product.

Please read this instruction manual carefully before using the instrument.

This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications.

These instruments are in compliance with CSA, UL and (EN 50081-1 and EN 50082-1) directives.

TABLE OF CONTENTS

PRELIMINARY EXAMINATION

3

GENERAL DESCRIPTION

3

FUNCTIONAL DESCRIPTION AND SPECIFICATIONS:

 

HI8417

5

HI8519

8

HI8520

10

HI8521

12

OPERATIONAL GUIDE

14

PH CALIBRATION

18

PROCEDURE FOR HI8417

18

PROCEDURE FOR HI8519

20

PROCEDURE FOR HI8520/HI8521

21

CAL. W/ MANUAL TEMP. COMPENSATION

22

PH VALUES AT VARIOUS TEMPERATURE

23

TEMPERATURE CALIBRATION

23

TROUBLESHOOTING GUIDE

24

ELECTRODE CONDITIONING & MAINTENANCE

25

TEMPERATURE-RESISTANCE CORRELATION

 

FOR HANNA PH SENSITIVE GLASS

28

ACCESSORIES

30

ELECTRODE APPLICATION REFERENCE GUIDE

37

WARRANTY

38

CE DECLARATION OF CONFORMITY

39

ISO 9000 Certified

Company since 1992

2

PRELIMINARY EXAMINATION

Remove the instrument from the packing material and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your dealer.

Note: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned in the original packing with the supplied accessories.

GENERAL DESCRIPTION

HI8417, HI8519, HI8520 and HI8521 Hanna Instruments benchtop pH meters are designed for simplicity of use.

They feature: large liquid crystal display (LCD), user friendly keyboard, error codes to guide the user in calibration and troubleshooting and storage of calibration data even after the unit is shut off.

HI8417, HI8519 and HI8521 can also measure ion concentration (ISE) and ORP in the mV range (with automatic range switching from ISE to ORP when reading exceeds ±399.9mV, HI8417 and HI8521 only).

All models, except for HI8519, are microprocessor-based and provide automatic or manual temperature compensation for both calibration and measurement.

These models meet the most stringent demands of the laboratory:

HI8417 pH/mV/°C meter automatically recognizes six standard calibration buffer values-three each for NBS (pH 4.01, 6.86, 9.18) technical standard calibration buffers (pH 4.01, 7.01, 10.01).

A recorder output of 0 to 2 Volts is available and the wide mV range and high resolution allow for ORP or ion-selective measurement. HI8417 comes supplied with HI1131B glass-body combination pH electrode, HI7669/2W temperature probe (with 1 meter / 3.3' cable), 12VDC adapter (HI710005 or HI710006) and a dust cover.

HI 8519 pH/mV meter includes: manual temperature compensation and front panel knobs for offset and slope calibration.

HI8519 comes supplied with HI1332B plastic-body combination pH electrode, 12VDC adapter (HI710005 or HI710006) and a dust cover.

3

HI8520 pH/°C meter automatically recognizes 3 standard calibration buffer values (4.01, 7.01 and 10.01 pH).

HI8520 comes supplied with HI1332B plastic-body combination pH electrode, HI7669/2W temperature probe (with 1 meter/ 3.3' cable), 12VDC adapter (HI710005 or HI710006) and a dust cover.

HI8521 pH/mV/°C meter automatically recognizes 3 standard calibration buffer values (4.01, 7.01 and 10.01 pH).

HI8521 comes supplied with HI1332B plastic-body combination pH electrode, HI7669/2W temperature probe (with 1 meter/ 3.3' cable), 12VDC adapter (HI710005 or HI710006) and a dust cover.

4

FUNCTIONAL DESCRIPTION &

SPECIFICATIONS OF HI8417

THE FRONT PANEL OF HI8417

°C

pH

HI 8417

 

BUFFER VALUES 20°C

°C

 

ON

d1

NBS

pH

6,88

CAL

 

 

t1

STD

pH

7,02

 

pH

 

d2

NBS

pH

4,00

 

mV

OFF

t2

STD

pH

4,00

°C

d3

NBS

pH

9,22

CON

 

 

t3

STD

pH

10,05

 

 

ON / OFF To turn the instrument ON and OFF. (Warning: Push the OFF key before unplugging instrument. If unplugged before turned off, the meter's memory will be erased).

CAL To enter the calibration mode. CON To confirm the calibration data.

°C / °C To set temperature for manual compensation.

pH/mV To select mV for mV, ORP or ISE measurements or pH for pH measurements.

Large LCD on the left

This LCD displays the temperature measured by the HI7669/2W probe or the value set manually with the keyboard. If the temperature has been set manually, the "°C" symbol will flash. When the temperature probe is connected the "°C" symbol stops flashing. When the temperature measured by the probe is out of range (-10 / +125°C), the "Err3" signal appears.

Large LCD on the right

This LCD displays the pH or mV reading. The corresponding pH or mV symbol will be displayed along with the measured value. An error signal ("Err1", "Err2", "Err4" or "Err5") will appear on this display to indicate when the range is exceeded or if the buffer during calibration is wrong.

5

THE REAR PANEL OF HI8417

 

 

 

 

TEMPERATURE

DC

POWER

COMBINED

 

 

 

PROBE

12V

to 20V

RECORDER

 

 

 

 

pH or ORP REFERENCE

 

 

 

+

ELECTRODES ELECTRODE RESET

0-2 VOLT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

2

3

4

5

6

1.BNC socket for pH and ORP electrode.

2.Reference Jack socket for reference electrode.

3.RESET button.

4.Recorder output connections.

5.Temp. Probe socket (for HI7669/2W).

6.DC Power Socket (for HI710005 or HI710006).

Electrode and probe connections

For combination pH or ORP electrodes (with internal reference) plug the electrode's BNC to the socket provided (#1).

For an electrode with a separate reference, connect the measuring electrode's BNC to the BNC socket (#1) and the reference electrode's jack to the socket provided (#2).

For temperature measurements and automatic temperature compensation connect the temperature probe to the °C socket (#5).

Recorder Output

HI8417 is provided with a recorder output from 0 to 2 volts. Connect the recorder output (#4) and select the pH or mV range by pressing the pH/mV key. The output signal is related to the values as follows:

pH output:

0.01 pH = 1.42857 mV e.g. 0.00 pH = 0 mV

7.00 pH = 1000 mV

14.00 pH = 2000 mV

mV output:

0.1 mV = 0.05 mV e.g. -1999 mV = 0 mV 0.0 mV = 1000 mV

+1999 mV = 2000 mV

The output impedance is 100 ohms. The maximum output shortcircuit current is 20 mA.

RESET button

The RESET button (#3) is used when the instrument displays erroneous messages due to strong electrical interference or when the

6

instrument's power supply is disconnected before the meter is switched off. Press the RESET button and restart the entire operation.

Power connection

Plug the 12VDC adapter (HI710005 or HI710006) into the DC socket (#6).

Note: Make sure the main line is protected by a fuse.

Note: HI8417 uses a rechargeable battery to retain the pH calibration data. It lasts approximately one month when fully charged. The battery is automatically charged any time the meter is plugged-in (either turned on or off).

SPECIFICATIONS

HI8417

Range

pH

0.00 to 14.00

 

ISE

0.0 to ±399.9mV

 

ORP

±1999mV

 

°C

-10.0 to 125.0

Resolution

pH

0.01

 

ISE

0.1mV

 

ORP

0.1/1mV

 

°C

0.1

Accuracy

pH

±0.01

(@ 20°C / 68°F)ISE

±0.2 mV

 

ORP

±1 mV

 

°C

±0.5

Typical EMC

pH

±0.01

Deviation

ISE/ORP

±0.2mV

 

°C

±0.7

Recorder

pH

0 to 2 V

Output

mV

0 to 2 V

Max. Current

20mA

Offset

 

±1 pH

Slope

 

from 85% to 105%

Calibration

 

Automatic 2 points with 5 memorized buffers

 

 

(pH 4.01, 6.86, 7.01, 9.18, 10.01)

Temperature

 

Manual from 0 to 100°C (32 to 212°F)

Compensation

Automatic from -10 to 125°C (14 to 257°F)

Electrode

 

HI1131B glass body combination refillable

 

 

pH electrode (included)

Temp. Probe

HI7669/2W Temperature Probe (included)

Input Impedance

1012 ohm

Power

 

Power socket for 12VDC

Environment

 

0 to 50°C (32 to 122°F);

 

 

95% RH (not condensing)

Dimensions

 

230Lx170Wx70H mm (9.1x6.7x2.7")

Shipping Weight

1.5 Kg (3.3 lb.)

 

 

 

7

Hanna Instruments HI8417, HI8519, HI8520, HI8521 User Manual

FUNCTIONAL DESCRIPTION &

SPECIFICATIONS OF HI8519

THE FRONT PANEL OF HI8519

pH

HI 8519

OFFSET

 

 

SLOPE

 

TEMPERATURE °C

 

OFFSET

 

 

SLOPE

 

TEMPERATURE °C

 

Temperature

Buffer Values (pH)

 

 

 

 

°C

°F

4.01

7.01

10.01

 

 

 

 

0

32

4.01

7.13

10.32

 

 

 

ON

10

50

4.00

7.07

10.18

pH

mV

°C

20

68

4.00

7.03

10.06

OFF

30

86

4.02

7.00

9.96

 

 

 

40

104

4.04

6.98

9.88

 

 

 

 

ON/OFF To switch the meter ON and OFF. pH To display pH values.

mV To display mV values.

°C To display temperature values set by the temperature knob (for pH compensation).

Knobs

OFFSET For adjustment of offset during pH calibration. SLOPE For adjustment of slope during pH calibration. TEMP °CFor setting temperature in °C.

THE REAR PANEL OF HI8519

COMBINED pH

DC

POWER

 

12V

to

20V

 

ORP

 

REFERENCE

 

+

 

ELECTRODES ELECTRODE

 

THIS METER IS NOT

 

 

 

 

PROTECTED BY FUSE

1

2

3

 

 

1.BNC socket for pH or ORP electrode.

2.Reference Jack socket for reference electrode.

3.DC Power Socket (for HI710005 or HI710006).

8

Electrode and probe connections

For combination pH or ORP electrodes (with internal reference) plug the electrode's BNC to the socket provided (#1).

For an electrode with a separate reference, connect the measuring electrode's BNC to the BNC socket (#1) and the reference electrode's jack to the socket provided (#2).

Power connection

Plug the 12VDC adapter (HI710005 or HI710006) into the DC socket (#3).

Note: Make sure the main line is protected by a fuse.

SPECIFICATIONS

HI 8519

Range

pH

0.00 to 14.00

 

mV

±1999

Resolution

pH

0.01

 

mV

1

Accuracy

pH

±0.01

(@20°C / 68°F) mV

±1

Typical EMC

pH

±0.06

Deviation

mV

±4

Calibration

 

Manual 2 points through

 

 

offset and slope knobs

Temperature

 

Manual from 0 to 100°C

Compensation

 

(32 to 212°F)

 

 

 

Electrode

 

HI 1332B plastic body, combination,

 

 

double junction refillable

 

 

pH electrode (included)

 

 

 

Input Impedance

 

1012 ohm

Power

 

12VDC power

Environment

 

0 to 50°C (32 to 122°F); 95% RH

Dimensions

 

230x170x70 mm (9.1x6.7x2.7")

Shipping Weight

 

1.35 Kg (3 lb.)

 

 

 

9

FUNCTIONAL DESCRIPTION &

SPECIFICATIONS OF HI8520

THE FRONT PANEL OF HI8520

ERROR SYMBOLS

 

 

CAL

 

Temperature

Buffer Values (pH)

ON

E1: pH out of range

 

 

 

 

°C

°F

4.01

7.01

10.01

OFF

E2: °C out of range

 

 

 

0

32

4.01

7.13

10.32

°C

 

 

 

 

 

E4: wrong pH 7

 

 

 

10

50

4.00

7.07

10.18

 

E5: wrong pH 4/10

 

 

CON

20

68

4.00

7.03

10.06

RANGE

 

 

30

86

4.02

7.00

9.96

 

 

 

 

40

104

4.04

6.98

9.88

 

ON / OFF

To switch instrument ON and OFF.

RANGE

To display pH or temperature measurements.

CAL/CON

CAL to enter the calibration procedure and CON to

 

confirm calibration data.

/

For setting temperature for manual compensation.

THE REAR PANEL OF HI8520

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TEMPERATURE

DC

POWER

 

 

 

 

 

 

 

PROBE

12V

to 20V

 

 

 

COMBINED

 

 

 

 

 

 

 

pH or ORP REFERENCE

 

 

 

+

 

 

 

ELECTRODES ELECTRODE RESET

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

2

3

4

5

1.BNC socket for pH electrode.

2.Reference Jack socket for reference electrode.

3.RESET button.

4.Temperature Probe socket (for HI7669/2W).

5.DC Power Socket (for HI710005 or HI710006).

Electrode and probes connections

For combination pH electrodes (with internal reference) plug the electrode's BNC to the socket provided (#1).

For an electrode with a separate reference, connect the measuring electrode's BNC to the BNC socket and the reference electrode's jack to the socket provided (#2).

For temperature measurements and automatic temperature compensation connect the temperature probe to the °C socket (#4).

RESET button

The RESET button (#3) is used when the instrument displays erroneous messages due to strong electrical interference or when the instrument's power supply is disconnected before the meter is switched

10

off. Press the RESET button and restart the entire operation.

Power connection

Plug the 12VDC adapter (HI710005 or HI710006) into the socket on the back of the meter (#5).

Note: Make sure the main line is protected by a fuse.

Note: HI8520 uses a rechargeable battery to retain the pH calibration data. It lasts approximately one month when fully charged. The battery is automatically charged any time the meter is plugged-in (either turned on or off).

SPECIFICATIONS

HI 8520

Range

pH

0.00 to 14.00

 

°C

0.0 to 100.0

Resolution

pH

0.01

 

 

°C

0.1

 

Accuracy

pH

±0.01

 

(@20°C / 68°F)

°C

±0.4

 

Typical EMC

pH

±0.02

 

Deviation

°C

±0.4

 

Calibration

 

Automatic 2 points with auto-buffer

 

 

recogntion

 

 

 

Temperature

 

Automatic or Manual

 

Compensation

 

From 0 to 100°C (32 to 212°F)

Electrode

 

HI 1332B plastic body, combination,

 

 

double junction refillable

 

 

pH electrode (included)

Temperature Probe

HI 7669/2W temp. probe (included)

Input Impedance

 

1012 ohm

Power

 

12VDC power

Environment

 

0 to 50°C (32 to 122°F); 95% RH

 

 

 

Dimensions

 

230x170x70 mm (9.1x6.7x2.7")

Shipping Weight

 

1.35 Kg (3 lb.)

 

 

 

 

11

FUNCTIONAL DESCRIPTION &

SPECIFICATIONS OF HI8521

THE FRONT PANEL OF HI8521

ERROR SYMBOLS

 

 

CAL

 

Temperature

Buffer Values (pH)

ON

E1: pH out of range

 

 

 

 

°C

°F

4.01

7.01

10.01

OFF

E2: °C out of range

 

 

 

0

32

4.01

7.13

10.32

°C

 

 

 

 

 

E3: mV out of range

 

 

 

10

50

4.00

7.07

10.18

 

E4: wrong pH 7

 

 

CON

20

68

4.00

7.03

10.06

RANGE

 

 

30

86

4.02

7.00

9.96

E5: wrong pH 4/10

 

 

 

40

104

4.04

6.98

9.88

 

ON / OFF To switch instrument ON and OFF.

RANGE To display pH / ORP / Temperature measurements. CAL/CON CAL to enter the calibration procedure and CON to

confirm calibration data.

/

 

 

 

For setting temperature for manual compensation.

THE REAR PANEL OF HI8521

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TEMPERATURE DC

POWER

 

 

 

COMBINED

 

 

 

PROBE

12V

to 20V

 

 

 

 

 

 

 

 

 

 

 

 

 

 

pH or ORP

REFERENCE

 

 

 

+

 

 

 

ELECTRODES ELECTRODE RESET

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

2

3

4

5

1.BNC socket for pH or ORP electrode.

2.Reference Jack socket for reference electrode.

3.RESET button.

4.Temperature Probe socket (for HI7669/2W).

5.DC Power Socket (for HI710005 or HI710006).

Electrode and probes connections

For combination pH or ORP electrodes (with internal reference) plug the electrode's BNC to the socket provided (#1).

For an electrode with a separate reference, connect the measuring electrode's BNC to the BNC socket and the reference electrode's jack to the socket provided (#2).

For temperature measurements and automatic temperature compensation connect the temperature probe to the °C socket (#4).

RESET button

The RESET button (#3) is used when the instrument displays erroneous messages due to strong electrical interference or when the

12

Loading...