Hanna Instruments HI 93734 User Manual

0 (0)
Hanna Instruments HI 93734 User Manual

Instruction Manual

HI 93734

Free & Total Chlorine HR

ISM

 

 

 

 

 

 

 

 

This Instrument is in

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

w w w . h a n n a i n s t . c o m

Compliance with CE Directives

Dear Customer,

Thank you for choosing a Hanna product. This manual will provide you with the necessary information for the correct operation of the meter. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.

This instrument is in compliance with directives.

Preliminary Examination

Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If there is any damage, notify your Dealer.

Each Ion Specific Meter is supplied complete with

9V Battery

Two Sample Cuvets and Caps

One Transport Cap

Note: Keep all packing material until the instrument has been observed to function correctly. Any defective item must be returned in its original packing.

Warranty

HI 93734 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions.

This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.

If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer

Service department and then send it with shipment costs prepaid.

When shipping any instrument, make sure it is properly packaged for complete protection.

To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.

All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.

Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.

General Description

The HI 93734 meter measures the Free and Total Chlorine (Cl2) content in water and wastewater.

The meter uses an exclusive positive-locking system to ensure that the cuvet is in the same position every time it is placed into the measurement cell.

The reagents are in powder and liquid form and are supplied in bottles and powder packets. The amount of reagent is precisely dosed to ensure maximum repeatability.

Display codes aid the user in routine operations.

The meters have an auto-shut off feature that will turn the instrument off after 10 minutes of non-use.

Specifications

SPECIFICATIONS

Range

Free Cl2 0.00 to 9.99 mg/L

 

Total Cl2 0.00 to 9.99 mg/L

Resolution

0.01 mg/L

Accuracy

±0.03 mg/L ±3% of reading

Typical EMC

±0.01 mg/L

Deviation

 

Light Source

Light Emitting Diode @ 555 nm

Method

Adaptation of the EPA recommended DPD

 

method 330.5. The reaction between

 

the chlorine and the DPD reagent causes

 

a pink tint in the sample

Light Detector

Silicon Photocell

Environment

0 to 50°C (32 to 122°F);

 

max 95% RH non-condensing

Battery Type/Life

1 x 9 volt/40 hours

Auto-Shut off

After 10' of non-use

Dimensions

180 x 83 x 46 mm (7.1 x 3.3 x 1.8")

Weight

290 g (10 oz.).

REQUIRED REAGENTS

Code

Description

Quantity

HI 93701-0

Free Chlorine Reagent

1 packet

HI 93734B-0

Free & Total Chlorine

5 mL

 

Reagent B

 

HI 93734C-0

Total Chlorine Reagent C

3 drops

 

(for Total Chlorine only)

 

REAGENT SETS

HI 93734-01 Reagents for 100 Free & Total Chlorine tests HI 93734-03 Reagents for 300 Free & Total Chlorine tests

Display Code Guide

This indicates that the meter is in a ready state and zeroing can be performed.

Sampling in Progress. This prompt appears each time the meter is performing a measurement.

This indicates that the meter is in a zeroed state and measurement can be performed.

A zero reading was not taken. Insert a sample before adding reagent and press ZERO.

Under range. A blinking "0.00" indicates

that the sample absorbs less light than the zero reference. Check the procedure and make sure you use the same cuvet for reference (zero) and measurement.

 

Over range. A flashing value higher than

 

the maximum concentration readable (see

 

specifications) indicates that the sample

 

absorbs too much light, meaning that the

 

concentration is too high. Dilute the

 

sample.

 

Light over range. The cuvet is not inserted

 

correctly and an excess ambient light is

 

reaching the detector. If the cover is

 

properly installed, then contact your dealer

 

or the nearest Hanna Customer Service

 

Center.

 

Light under range. The zero sample is too

 

dark for proper zeroing. If this is not the

 

case, contact your dealer or the nearest

 

Hanna Customer Service Center.

V

The "V" indicates that the battery voltage

 

is getting low and the battery needs to be

 

replaced.

 

This indicates that the battery is dead and

 

must be replaced.

 

Note: once this indication is displayed,

 

the meter will lockup. Change the battery

 

to restart.

Loading...
+ 1 hidden pages