Emerson 396VP, 396, 398VP, 398, 397 User Manual

0 (0)
pH/ORP Sensors
Instruction Manual
PN 51-396/397/398/rev.L November 2010
MODEL 396 and MODEL 398
MODEL 396VP and MODEL 398VP
MODEL 397
CAUTION
SENSOR/PROCESS
APPLICATION COMPATIBILITY
The wetted sensor materials may not be com­patible with process com position and operat­ing conditions. Application compat ibility is entirely the responsibility of the user.
DANGER
HAZARDOUS AREA INSTALLATION
This sensor is not Intrinsically Safe. or Explosion Proof. Installations near flammable liquids or in hazardous area locations must be carefully evaluated by qualified on site safety personnel.
To secure and maintain an intrinsically safe installation, an appropriate transmitter/safety barrier/sensor combi nation must be used. The installation system must be in accordance with the governing approval agency (FM, CSA or BASEEFA/CENELEC) hazardous area classifi­cation requirements. Consult your analyzer/ transmitter instruc tion manual for details.
Proper installation, operation and servicing of this sensor in a Hazardous Area Instal lation is entirely the responsibility of the user.
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures, and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use, and main­tain them to ensure they continue to operate within their normal specifica­tions. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; proper­ty damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the prod­uct. If this Instruction Manual is not the correct manual, telephone 1-800­654-7768 and the requested manual will be provided. Save this Instruction Manual for future reference.
• If you do not understand any of the instructions, contact your Rosemount representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
• Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and pro­cedures can affect the product’s performance and place the safe opera­tion of your process at risk. Look alike substitutions may result in fire, electrical hazards, or improper operation.
About This Document
This manual contains instructions for installation and operation of the Models 396, 396VP, 397, 398, and 398VP TUpH pH/ORP Sensors. The following list provides notes concerning all revisions of this document.
Rev. Level
Date Notes
0 3/99 This is the initial release of the product manual. The manual has been reformatted to reflect the
Emerson documentation style and updated to reflect any changes in the product offering.
A 12/01 Revised wiring diagram on page 24.
B 2/02 Added info to page 1.
C 6/02 Updated multiple drawings.
D 8/02 Added drawing #40105549, rev. D.
E 10/02 Revised drawing #40039601, rev. J, on page 10.
F 4/03 Revised Model 397 specs on page 2, and revised drawing on page 27.
G 8/03 Added Silcore information.
H 3/04 Added Xmt wiring drawings.
I 10/04 Added 5081 wiring drawing and updated 1055 wiring.
J 2/06 Changed drawing 40039603 rev. C, on page 13. Added a note on page 15.
K 1/07 Miscellaneous revisions
L 11/10 Removed mention of patents and updated dnv logo
i
MODEL 396/396VP/397/398/398VP TABLE OF CONTENTS
TUpH MODELS 396/396VP/397/398/398VP
pH/ORP SENSORS
TABLE OF CONTENTS
Section Title Page
1.0 DESCRIPTION AND SPECIFICATIONS ......................................................... 1
1.1 TUpH™ Features and Applications.................................................................. 1
1.2 Performance and Physical Specifications ........................................................ 3
1.3 Ordering Information ........................................................................................ 4
2.0 INSTALLATION ................................................................................................ 9
2.1 Unpacking and Inspection................................................................................ 9
2.2 Mounting .......................................................................................................... 9
2.3 Electrical Installation ....................................................................................... 15
3.0 START UP AND CALIBRATION....................................................................... 31
3.1 Sensor Preparation .......................................................................................... 31
3.2 Calibration Using Buffer Solutions or Grab Samples ....................................... 31
4.0 MAINTENANCE ............................................................................................... 32
4.1 Electrode Cleaning........................................................................................... 32
4.2 Automatic Temperature Compensator ............................................................. 32
5.0 TROUBLESHOOTING ..................................................................................... 33
6.0 RETURN OF MATERIAL ................................................................................. 34
LIST OF TABLES
Number Title Page
4-1 R0and R1Values for Temperature Compensation Elements.......................... 22
4-2 Temperature vs Resistance of Auto T.C. Elements .......................................... 22
5-1 Troubleshooting ............................................................................................... 23
MODEL 396/396VP/397/398/398VP TABLE OF CONTENTS
LIST OF FIGURES
Number Title Page
1-1 Cross Section Diagram of the TUpH Reference Technology............. 2
2-1 Flow Through/Insertion Installation for Models 396, 396VP, 398, & 398VP .............. 10
2-2 Dimensional Drawing For Models 396 & 398 .......................................................... 11
2-3 Dimensional Drawing For Models 396VP & 398VP ................................................ 12
2-4 Submersion Installations For Models 396, 396VP, 398, & 398VP ........................... 13
2-5 Dimensional Drawings For Model 397 in the Quik-Loc Unit ................................... 14
2-6 Recommended Flow Through and Insertion Installation For Model 397 ................ 15
2-7 Models 396/397/398 Standard Wiring option used with Models 1054A, ...............
1054B, 2054, and Preamps PN 22698-02 & 22698-03.......................................... 15
2-8 Models 396/397/398 Option -62 for wiring to Models 1054, 1181, 2081, 54, ........
81, 3081, 4081, SoluComp, and 2700 Preamp (PN 23054-03) and remote .........
J-box PN 23555-00................................................................................................. 15
2-9 Wiring Models 396-54 (50), 397-54 (50), & 398-54 (50) to J-Box .........................
(PN 23309-04 (03)) & Remote Preamplifier ........................................................... 16
2-10 Wiring Models 396-54, 397-54, & 398-54 (PT-100-RTD) to J-Box (PN 23555-00)
& Remote Preamplifier ........................................................................................... 17
2-11 Wiring Models 396/397/398 to Model 1181pH........................................................ 18
2-12 Wiring Models 396-54, 397-54, & 398-54 to Models 1054A pH, 1054B pH, .........
& 2054pH-54 .......................................................................................................... 19
2-13 Wiring Models 396-54, 397-54, & 398-54 to Models 2081pH-05 & 1054pH-54........ 20
2-14 Wiring Models 396-54, 397-54, & 398-54 to Models 54, 81, 3081, & 4081............ 21
2-15 Wiring Models 396-54, 397-54, & 398-54 to SoluComp Model SCL-P-014 ........... 22
2-16 Wiring Models 396-54, 397-54, & 398-54 to Model 2700 Preamp ......................... 23
2-17 Wiring Models 396-( )-54-62-(71), 397-02-10-54-62, & 398-( )-54-62 to ...............
Model 1055-01-10-22-32 ........................................................................................ 23
2-18 Wiring Models 396-54-62, 397-02-54-62, & 398-54-62 to Model 1055-22-32........ 23
2-19 Wire Functions and Pin Connections for Variopol Connector Cable ...................... 24
2-20 Wiring Models 396VP/398VP to Model 81 thru a remote J-box ............................. 24
2-21 Wiring Models 396VP/398VP to Model 81 ............................................................. 24
2-22 Wiring Models 396VP/398VP to Model 1181.......................................................... 24
2-23 Wiring Models 396VP/398VP to Model 1181, 1050/1060, 1003/1023 thru a ........
remote J-box........................................................................................................... 25
2-24 Wiring Models 396VP/398VP to Model 2081 ......................................................... 25
2-25 Wiring Models 396VP/398VP to Model 3081 & 4081 ............................................. 25
2-26 Wiring Models 396VP/398VP to Model 2081 thru a remote J-box ......................... 25
2-27 Wiring Models 396VP/398VP to Model 3081 & 4081 thru a remote J-box............. 25
2-28 Wiring Models 396VP/398VP to Model 1054 ......................................................... 26
2-29 Wiring Models 396VP/398VP to Model 1054A/B & 2054 ...................................... 26
2-30 Wiring Models 396VP/398VP to Model 1054 thru a remote J-box ......................... 26
2-31 Wiring Models 396VP/398VP to Model 1054A/B & 2054 thru a remote J-box....... 26
2-32 Wiring Models 396VP/398VP to Model 54/54e thru a remote J-box ...................... 27
2-33 Wiring Models 396VP/398VP to Model 54/54e ...................................................... 27
2-34 Wiring Models 396VP/398VP to Model SCL-(P/Q)................................................. 27
2-35 Wiring Models 396VP/398VP to Model 2700 ......................................................... 27
2-36 Wiring Models 396VP/398VP to Model 1055-22-32 ............................................... 28
2-37 Wiring Models 396/397/398 to Model Xmt.............................................................. 29
2-38 Wiring Models 396VP/398VP to Model Xmt ........................................................... 29
2-39 Wiring Models 396/397/398 to Model 5081-P ........................................................ 30
2-40 Wiring Models 396-54-62/398-54-62 to Model 1055-01-10-22-32 ......................... 30
ii
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.1 TUpH™ FEATURES AND APPLICATIONS.
Rosemount Analytical has achieved a new industry standard for the life expectancy of pH sensors with the TUpH Sensors. The low maintenance, disposable Models 396, 396VP, 397, 398, and 398VP TUpH Sensors offer long life and high performance in the measurement of pH in aqueous solutions in pipelines, open tanks, or ponds. These TUpH sensors feature a large area reference junction for maximum resistance to process coatings, generally found in dirty, high solid applications. In addition, the secondary helical refer­ence pathway serves as added protection from poi­soning ions. The simplified construction, designed with user convenience in mind, does not require elec­trolyte replenishment or any component replacement. All models feature a large glass bulb for increased resistance to the effects of aging for longer life. All models are available only without an integral pream­plifier. The preamplifier must be in a remote location or integral to the analyzer/transmitter. Models with a “VP” after the model number indicate the variopol connector is used on the sensor in place of the cable. In this case, a separate cable with a mating VP con­nector must be used. All models are compatible with all Rosemount Analytical and most manufacturers instruments.
The entire line of TUpH model sensors now incor­porate the new SILCORE technology contaminant barrier. This triple-seal barrier prevents moisture and
material impurities from migrating to the pH sensor’s reference electrode’s metal lead wire. By preventing these contaminants from compromising the integrity of the pH measurement, sensor life is increased, espe­cially at higher temperatures where increased migra­tions occur. In addition, the SILCORE
1
technology pro­vides added protection against sensor failure due to vibrations and shock by transferring damaging energy away from the glass-to-metal seal.
1.1.1 Model 396 and 396VP TUpH Features and Applications.
The Models 396 and 396VP TUpH sensors are construct­ed of polypropylene and stain­less steel and is completely sealed by EP (ethylene propy­lene) to eliminate process intrusion. It is specifically designed for improved life in harsh, dirty applications such as lime slurry waste treatment and paper machine headbox and pigment/dye applications where large quantities of suspended solids are present. Installation is easily achieved through a wide variety of
mounting configurations. The Model 396 features an optional front or rear 1 in. MNPT process connection for insertion, submersion, or flow through applications.
1.1.2 Model 397 TUpH and Quik-Loc Features and Applications. The Model 397
is housed in a highly chemical resistant polypropylene body and completely sealed with EP to eliminate process intru­sion. The Model 397 body is specifically designed for use with the Quik-Loc Kit which consists of an adapter and coupler. The PEEK (polyetheretherketone) adapter enables the Model 397 sensor to fit into a 1 in. MNPT Dixon coupler for quick and easy removal without sensor cable twisting. The 316 stainless steel DIxon coupler is sealed with EP and features locking arms. The Quik-Loc Kit is not recom­mended for use in processes with hazardous, corro­sive, or strong oxidizing chemicals due to a risk of spray and bodily hazards.
1.1.3 Model 398 and 398VP TUpH Features and Applications.
The chemical-resistant con­struction of Tefzel, titanium, and the TUpH reference junc­tion make the Models 398 and 398VP the ideal sensors for measuring pH in harsh process liquids. Use Model 398 or 398VP to measure pH in sour water strippers, in pulp bleaching towers that use chlorine dioxide, and in process streams containing a variety of organic sol­vents.
Models 398 and 398VP use the highest quality mate­rials to provide superior chemical resistance. The sen­sors are housed in a titanium tube and features an optional 1 inch MNPT process connector for insertion, submersion, or flow-through applications. The molded Tefzel TUpH construction is offered with a choice of seals (Viton, EPDM, or Kalrez). Combining high qual­ity materials with the TUpH reference technology and A
CCUGLASS
1
pH bulb allows for ultimate chemical resistance and makes Models 398 and 398VP the perfect choice for measuring pH in harsh, demanding processes.
1
ACCUGLASS and Silcore are trademarks of Rosemount Analytical
MODEL 396/396VP/397/398/398VP SECTION 1.0
DESCRIPTIONS AND SPECIFICATIONS
1
2
MODEL 396/396VP/397/398/398VP SECTION 1.0
DESCRIPTIONS AND SPECIFICATIONS
FIGURE 1-1. Cross Section Diagram of the TUpH Reference Technology
All TUpH sensors are designed with a large area reference junction, helical reference pathway, and an AccuGlass pH
glass bulb. This sensor technology ensures superior performance while only requiring minimal maintenance.
3
MODEL 396/396VP/397/398/398VP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.2 PERFORMANCE & PHYSICAL SPECIFICATIONS
SPECIFICATIONS
MODELS 396 &396VP MODELS 398 & 398VP MODEL 397 QUIK-LOC KIT
Measurements pH: 0-14 pH: 0-14 pH: 0-14
and Ranges ORP: -1500 to 1500 mv
Available pH GPHT hemi bulb or GPHT hemi bulb or GPHT hemi bulb
ACCUGLASSTypes GPLR flat bulb GPLR flat bulb
Wetted Materials 316 SST, Polypropylene, Titanium, Tefzel, glass, Polypropylene, EP, glass 316 SST, EP, PEEK
EPDM, glass choice of Kalrez, Viton, or
EPDM (platinum: ORP only)
Process Connection None, use 1 in. NPT None, use 1 in. NPT None, must use Quik-Loc 1-in. MNPT
process connector, process connector, kit which includes 1 in.
PN 23166-00 or 23166-01 PN 23166-00 or 23166-01 MNPT process connection
(sold separately) (sold separately)
Temperature Range 0-100°C (32-212°F) 0-100°C (32-212°F) 0-100°C (32-212°F)
Pressure Range- 100-1136 kPa abs 100-1825 kPa abs 100-790 kPa abs
Hemi bulb (0-150 psig) (0-250 psig) (0-100 psig)
Pressure Range- 100-790 kPa abs 100-790 kPa abs
Flat bulb (0-100 psig) (0-100 psig)
Minimum Conductivity 75 µS/cm, nominal 75 µS/cm, nominal 75 µS/cm, nominal
Preamplifier Options Remote Remote Remote
Weight/Shipping Weight 0.45 kg/0.9 kg (1 lb/2 lb) 0.45 kg/0.9 kg (1 lb/2 lb) 0.45 kg/0.9 kg (1 lb/2 lb) 0.45 kg/0.9 kg (1 lb/2 lb)
PERCENT LINEARITY
396, 396VP, 397, 398, 398VP 396, 396VP, 397, 398, 398VP
pH Range GPHT Hemi GPLR Hemi
0-2 pH 94% 93%
2-12 pH 99% 98%
12-13 pH 97% 95%
13-14 pH 92%
4
MODEL 396/396VP/397/398/398VP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.3 ORDERING INFORMATION
MODEL
396 TUpH INSERTION/SUBMERSION STAINLESS STEEL pH SENSOR
CODE ANALYZER/TC COMPATIBILITY (Required Selection)
50 For use with Models 1181, 1050, 1003 (3 K TC)
54 For use with Models 54/e, 1054A/B, 1055, 2700, SCL-P, 81, 2081, 3081, and 4081 (Pt-100 RTD)
CODE OPTIONAL ITEMS
62 Cable without BNC for wiring to 4081/3081/81/54 analyzers
71 Flat bulb, GPLR
396 - 54 EXAMPLE
The Model 396 pH Sensor features your selection of glass electrode type, either the
standard hemi bulb or the optional flat glass electrode combined with the coating resistant, polypropylene reference with gel filled electrolyte. The sensor is housed in a stainless steel body and is used with a 1 in. MNPT threaded process connector suitable for insertion, submersion, or flow through installations. The Model 396 is available without an integral preamplifier only and 15 ft of cable. Automatic temperature compensation is standard 3 K Balco or PT-100 RTD.
MODEL
396VP TUpH INSERTION/SUBMERSION STAINLESS STEEL AND POLYPROPYLENE pH SENSOR (GPHT hemi bulb)
CODE ANALYZER/TC COMPATIBILITY (Required Selection)
50 For use with Models 1181, 1050, 1003 (3 K TC)
54 For use with Models 54/e, 1054A/B, 1055, 2700, SCL-P, 81, 2081, 3081, and 4081 (Pt-100 RTD)
CODE OPTIONAL OPTIONS
71 GPLR Flat bulb
396VP - 54 EXAMPLE
The Model 396VP insertion/submersion sensor is the same as Model 396 except,
instead of a cable on the back end of the sensor, it features a Variopol connector. The watertight Variopol sensor-to-cable connector uses a mating connector cable (ordered separately).
MODEL 396/396VP/397/398/398VP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
The Model 398 Sensor, housed in a titanium tube and made with a Tefzel reference junction, can be used with a 1 inch MNPT process connector (purchased separately). The sensor is available with a hemi bulb glass pH electrode or a platinum and glass ORP electrode and with Pt100 or 3K temperature compensation. The 398 is provided with a standard 15 ft cable. A preamplifier must be used if the sensor is more than 15 ft from the analyzer/transmitter. Junction box kits with preamplifiers must be ordered sep­arately. Process connector must also be ordered separately.
CODE MEASURING ELECTRODE TYPE (Required Selection)
10 GPHT hemi glass, General Purpose High Temperature (0-14 pH)
12 ORP
13 Flat Bulb, GPLR
CODE ANALYZER/TC COMPATIBILITY (Required Selection)
50 For Models 1181 (3K TC) [no T.C. if ordered with option -12 (ORP)]
54 For Models 54/e, 1054A/B, 1055, 2700, 81, 2081, 3081, 4081, SCL-P, and Solu Cube (Pt-100 RTD)
CODE O-RING MATERIAL (Required Selection)
30 EPDM
31 Viton
32 Kalrez (recommended for applications with Chlorine Dioxide)
CODE OPTIONAL SELECTIONS (see drawings on page 4)
62 No BNC (Not Valid w/Option 50) for wiring directly to 54/e, 1055, 81, 3081, and 4081 Transmitter/Analyzers
MODEL
398 TUpH INSERTION/SUBMERSION TITANIUM pH SENSOR
398 - 10 - 32 - 54 EXAMPLE
The Model 398VP insertion/submersion sensor is the same as Model 398, except
the cable is replaced with a Variopol connector. The watertight Variopol sensor-to­cable connector uses the mating connector cable (ordered separately).
CODE ANALYZER/TC COMPATIBILITY (Required Selection)
50 For Models 1181 (3K TC) [no T.C. if ordered with option -12 (ORP)]
54 For Models 54/e, 1054A/B, 1055, 2700, 81, 2081, 3081, 4081, SCL-P, and Solu Cube (Pt-100 RTD)
CODE MEASURING ELECTRODE TYPE (Required Selection)
10 GPHT hemi glass, General Purpose High Temperature (0-14 pH)
12 ORP
13 Flat
CODE O-RING MATERIAL (Required Selection)
30 EPDM
31 Viton
32 Kalrez (recommended for applications with Chlorine Dioxide)
MODEL
398VP TUpH INSERTION/SUBMERSION TITANIUM AND TEFZEL pH/ORP SENSOR
398VP - 10 - 32 - 54 EXAMPLE
5
6
MODEL 396/396VP/397/398/398VP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
The TUpH Model 397 pH Sensor is housed in a polypropylene body and is designed
to be used with the Quik-Loc Kit. The sensor includes a large general purpose (GPHT) glass pH electrode and a large area polypropylene reference junction with gel filled reference electrolyte. The Model 397 is available without an integral preamplifier only and 15 ft (4.6 m) of cable.
MODEL
397 TUpH pH SENSOR
CODE PREAMPLIFIER/CABLE (Required Selection)
02 Without integral preamplifier, 15 ft (4.6 m) cable
CODE MEASURING ELECTRODE TYPE (Required Selection)
10 GPHT, General Purpose glass (0-14 pH)
CODE ANALYZER/TC COMPATIBILITY (Required Selection)
50 For Models 1181, 1050, 1003 (3 K Balco TC)
54 For Models 1054, 1055, 54e, 81, 5081, 3081, 4081, 2081; Xmt (PT-100 RTD)
CODE OPTIONAL OPTIONS
62 Cable without BNC for wiring to 54/e, 1055, 81, 3081, 4081 analyzers, Xmt
1. Quik-Loc Mounting (required for all first time installations)
Choose one: PN
23757-00, Quik-Loc Kit: for use in 1 in. tees; insertion depth 1.4 in. (35 mm)
PN
23757-01, Quik-Loc Kit: for use in 1-1/2 in. and 2 in. tees; insertion depth 2 in. (50 mm)
2. Remote Junction Boxes (optional, recommended for sensor to analyzer distances of more than 15 ft)
Choose one: PN 23555-00 includes preamplifier for Models 54, 81, 3081, 4081
PN 23309-03 and PN 22698-02 plug-in preamplifier for Model 1181Analyzer
PN 23309-04 and PN 22698-03 plug-in preamplifier for Models 1054 series, 2054, and 2081 Analyzers
PN 23054-03 includes preamplifier for Solu Cube Model 2700
3. Extension cables (used with remote junction boxes)
Choose one: PN 23646-01, 11 conductor, shielded, prepped
PN 9200273, 11 conductor, shielded, unprepped
4. Wiring Aids
Choose one: PN 9120531 BNC splitter, used in place of option -62 or BNC coax prepping for connections to
junction box (PN 23555-00); Models 1181, 1054, 2081, 54, 81, 3081, 4081; Solu Comp (SCL-P-014);
and Model 2700 Preamplifier (PN 23054-03)
NONE: No wiring aids required for connections to Models 1054A, 1054B, 2054, or preamplifiers
PN 22698-02 or 22698-03
FOR FIRST TIME 397/QUIK-LOC INSTALLATIONS, USING THE FOLLOWING GUIDE IS RECOMMENDED:
7
MODEL 396/396VP/397/398/398VP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1. Process Connector Accessories (required for all first time installations with 1-inch process connection threads)
Choose one: PN 23166-00, 316 SST, 1 in. x 1 in. NPT process connector, with EPDM o-ring
PN 23166-01, Titanium, 1 in. x 1 in. NPT process connector, with EPDM o-ring
PN 9510066, Nylon, 1 in. x 1 in. NPT process connector (submersion only)
Choose one (optional process connector o-rings)
PN 9550220, Kalrez o-ring, 2-214
PN 9550099, Viton o-ring, 2-214
2. Variopol Cable (required for all first time installations) of Models 396VP and 398VP
Choose one: PN 23645-06, 15 ft cable with mating VP connector, prepped with BNC on analyzer end
PN 23645-07, 15 ft cable with mating VP connector, prepped without BNC on analyzer end*
3. Mounting Accessories (optional)
Choose one: PN 915240-03 PVC flow through tee, 3/4 in. NPT process connection
PN 915240-04 PVC flow through tee, 1 in. NPT process connection
PN 915240-05 PVC flow through tee, 1 1/2 in. NPT process connection
PN 11275-01 Sensor handrail mounting assembly
PN 2002011 1-1/2 in. CPVC Tee with 1 in. FNPT connection
PN 23728-00 Low Flow Cell, acrylic
4. Remote Junction Boxes (optional, recommended for sensor to analyzer distances of more than 15 ft)
Choose one: PN 23555-00 includes preamplifier for Models 54/e, 1055, 81, 3081, 4081
PN 23309-03 and PN 22698-02 plug-in preamplifier for Model 1181Analyzer
PN 23309-04 and PN 22698-03 plug-in preamplifier for Models 1054 series, 2054, 2081 Analyzers
PN 23054-03 includes preamplifier for Solu Cube Model 2700
5. Extension cables (used with remote junction boxes)
Choose one: PN 23646-01, 11 conductor, shielded, prepped
PN 9200273, 11 conductor, shielded, unprepped
6. Wiring Aids
Choose one: PN 9120531 BNC splitter, used in place of option -62 or BNC coax prepping for connections to junction box
(PN 23555-00); Models 54, 1055, 1054, 81,1181, 2081, 3081, 4081; SoluComp (SCL-P-014);
and Model 2700 Preamplifier (PN 23054-03)
NONE: No wiring aids required for connections to preamplifier PN 22698-02, preamplifier PN 22698-03,
or Models 1054A, 1054B, and 2054
FOR FIRST TIME 396, 396VP, 398, AND 398VP INSTALLATIONS, ROSEMOUNT ANALYTICAL RECOMMENDS USING THE FOLLOWING GUIDE:
*Used for connections to Models 54/e, 1055, 1054, 81, 1181, 2081, 3081, 4081, and remote junction box PN 23555-00.
8
MODEL 396/396VP/397/398/398VP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
ADDITIONAL ACCESSORIES FOR MODEL 396, 396VP, 398 AND 398VP TUpH SENSORS
PART NUMBER DESCRIPTION
33046-00 Ferrule 1 in., split 316 SS
33211-00 Adapter retrofit for PN 915240-04
9550167 EPDM O-ring for process connector (PN 23166-00)
9310100 Ferrule, 1 in. Teflon
1
9310096 Nut, Swage, 1 in. stainless steel
9550167 O-ring, 2-214 EPDM
1
Teflon is a registered trademark of E.I. du Pont de Nemours & Co.
ADDITIONAL ACCESSORIES FOR MODEL 397 TUpH SENSOR AND THE QUIK-LOC KIT
PART NUMBER DESCRIPTION
23753-00 PEEK replacement adapter for Quik-Loc kit (PN 23757-01)
23753-01 PEEK replacement adapter for Quik-Loc kit (PN 23757-00)
9160441 1 in. MNPT 316 stainless steel coupler
9160442 1 in. Plug, stainless steel
9160447 1 in. EP Gasket for coupler
2002011 1-1/2 in. CPVC Tee with 1 in. FNPT Connection
ADDITIONAL ACCESSORIES USED FOR MODELS 396, 396VP, 397, 398, AND 398VP TUpH SENSORS
PART NUMBER DESCRIPTION
22698-00 Preamplifier plug-in for J-box, 1003 compatible
22698-02 Preamplifier plug-in for J-box, 1181/1050 compatible
22698-03 Preamplifier plug-in for J-box, 1054A/B, 2054,2081 compatible
22743-01 Pt 100 preamplifier, 1181 compatible
22744-01 3K Preamplifier, 1181 compatible
23054-03 Remote J-box with preamplifier for Solu Cube Model 2700
23309-03 Remote Junction box, for preamplifier Model 1181
23309-04 Remote Junction box, for preamplifier Model 1054 Series, 2054, 2081
23555-00 Remote Junction box, with preamplifier, 54/e, 81, 3081, 4081 compatible
23557-00 Preamplifier for Junction box for Models 54/e, 1055, 81, 3081, 4081
23646-01 Extension cable 11 conductor, shielded, prepped
9200273 Extension cable11 conductor, shielded, unprepped
9210012 Buffer solution, 4.01 pH, 16 oz.
9210013 Buffer solution, 6.86 pH, 16 oz.
9210014 Buffer solution, 9.18 pH, 16 oz.
2001492 Stainless Steel Tag, Specify Marking
9200254 Cable, 4 cond., 22 AWG, 2 shielded pair