ULTRASONIC NEBULIZER
MODEL CUN60
Handy design - with built-in detachable battery pack; compact and light-
weight.
Easy to use - uses water as ultrasonic conducting medium, no gel re-
quired.
Silent operation - thanks to the ultrasonic vibration technology.
3-speed nebulization rate control - provides optimal treat-ment for all
kinds of medication with different density.
Built-in timer - automatically shuts off after 10-minute use.
Electromagnetic compatibility - complies with EN60601-1-2 requirement
as specified in EC directive 89/336/EEC.
FEATURES
English
Español
Português
Deutch Français
IMPORTANT SAFEGUARDS
READ ALL INSTRUCTIONS CAREFULLY BEFORE USE
PREPARATION
Use the equipment only for its intended use as de-
scribed in this manual. Do not use attachments not
recommended by the manufacturer.
Use the equipment with medications only under the
instruction of your physician.
Do not use the equipment if it has any damaged parts,
or it has fallen into water. Return the product to a ser-
vice center for examination and repair.
Always unplug this product before filling and after use.
To use the equipment for the first time, or after stor-
ing it for an extended period, be sure to clean and
sterilize all necessary parts as described on page 8.
Do not handle the power cord with wet hands or im-
merse the AC adapter. An electrical shock may result.
The medication cup, mouth nozzle and air filter should
be checked before each use. Replace with a new one
if they are dirty or worn out.
Do not fill medication cup over 8 ml or the nebuliza-
tion will be very weak.
OPERATION
Close supervision is required when this equipment is
used by children or invalids.
Stop using the equipment immediately if any abnor-
mality occurs.
Keep your eyes away from the output of medication
mist.
Do not use the equipment where flammable gas is
being used.
Never operate the equipment with empty water tank
or medication cup.
Do not use solutions other than water for the tank. It
may damage the transducer. Change the water after
each use.
Do not tilt or shake when operation.
Never open the clear cover during operation. It may
cause a severe injury.
Do not use the equipment for more than 30 minutes.
Do not reuse within 30 minutes.
Some medications may not have good atomization.
Check with the physician or manufacturer of medica-
tion.
CLEANING
Never immerse the equipment in water. It may dam-
age the unit.
Clean all necessary parts after each use, as described
on page 8.
STORAGE
Do not store the unit under direct sunlight, high tem-
perature or humidity.
Keep the unit out of reach of small children.
Always keep the unit unplugged while not in use.
Manufacturer:
Name: Vega Technologies Inc.
Address: 11F-13, 100 Chang-Chan Road,
Taipei, Taiwan
English
Español
Português
Deutch Français
PARTS DESCRIPTION
Mouth nozzle
Filter cover
Air filter
Top cover
Medication cup
Fan
Water reservoir
Low-battery
indication lamp
Main body
Nebulization rate
Indication lamps
ON/OFF button
Nebulization rate
control button
Rechargeable
Battery pack